Figure 2-2: Rosemount FCL-02
A. pH sensor
B. Chlorine sensor
C. Inlet
D. Drain
A
¼
-in. OD tubing compression fitting is provided for the sample inlet. If desired, you can
remove the compression fitting and replace it with a barbed fitting. The fitting screws into
a
¼
-in. FNPT check valve. The check valve prevents the sensor flow cell from going dry if
sample flow is lost.
The sample drains through a
¾
-in. barbed fitting.
1. Attach a piece of soft tubing to the fitting and allow the waste to drain to open
atmosphere.
Important
Do not restrict the drain line.
Install
Reference Manual
May 2019
00809-0100-3412
14
Emerson.com/Rosemount