2.3.2
AC power safety information
NOTICE
This section is in accordance with IEC 60079-0:2011 Clause 30.
This section must not be changed, amended, or removed.
WARNING
ELECTRIC SHOCK
Failure to follow instructions may cause personal injury or death.
Only qualified personnel, taking into account all applicable standards and legislative
requirements, should install the analyzer and connect the power and signal cables.
Instruments providing screw terminals for electrical connections may require working
near live parts.
The analyzer does not have power switches and is operable when connected to power.
A customer-supplied power switch or circuit breaker (complying with IEC 60947-1/ -3)
must be in the building installation. The switch has to be installed near the analyzer,
must be easily accessible, and has to be assigned as a disconnector for the analyzer.
Disconnect instruments with screw terminals from power when working at power
terminals (pull power plug or operate power switch / circuit breaker in building
installation).
The analyzer provides a protective earth terminal. To prevent electrical shock hazards,
connect the instruments to a protective earth. The instruments must be connected to
power with a three wire power cable with earth conductor. Any interruption of the
earth connector inside or outside the instrument or disconnecting the earth terminal
may cause potential electrical shock hazard.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Failure to observe this warning could cause an explosion or potentially hazardous situation
which, if not avoided, may cause death or personal injury.
Only enable the purge controller bypass function during setup or maintenance and only
when the area is known to be non-hazardous.
The customer supplied circuit breaker must be in accordance with ATEX / IECEx / North
American protection concepts. The main power isolator controls the application of
electrical power to the analyzer.
Electrical protection for the instrumentation circuitry of the analyzer is provided by fuses
F1 and F2 located inside the analyzer.
Install
Reference Manual
September 2019
00809-0100-5800
42
Emerson.com/Rosemount
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...