The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard; see the instructions for guidance.
•
Ensure that the flow of process gas is limited to a mazimum flow rate of 1.6 gallons (6
liters) per minute.
•
Only use the analyzer with process gases which are in gas groups B, C or D (Divisions) or
IIB + H
2
(Zones) and do not use oxygen or any other oxidizer in concentrations greater
than that found in normal air.
•
This assessment does not cover reliable function, performance, or other properties of
the equipment not related to safety.
•
Install the analyzer using wire no larger than 1mm
2
(18 AWG).
•
Only manufacturer trained personnel should install the analyzer.
•
If at any time there is a conflict between the system safety provisions and any relevant
local (national or regional) requirements, local requirements always take precedence.
•
Do not use the analyzer with flammable liquids.
1.6
Qualified personnel
This manual provides installation, operation, and maintenance personnel with the level of
knowledge required to safely start, operate, and switch off the analyzer.
Only technically qualified personnel in the field of analysis and control who are familiar
with this manual and have been specially trained on the analyzer should install, operate,
switch off, and service the analyzer. Only qualified and trained persons have the required
specific knowledge to correctly interpret the general safety information, warnings, and
procedures given in this manual and apply them to this particular application. Emerson or
its distribution partners can provide this training on request.
Knowledge of the safety information within this manual and its technically correct
implementation are prerequisites for danger-free installation, operation, and maintenance
of the analyzer.
1.7
Software version
The analyzer includes software that is used to control the operation of the analyzer. This
manual describes the software version as: 5.12.7 minimum .
1.8
Gas detection
The analyzer is highly configurable in the gases that can be detected and their range of
concentrations.
Reference Manual
Plan
00809-0100-5800
September 2019
Rosemount CT5800
27
Summary of Contents for Rosemount CT5800
Page 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Page 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Page 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Page 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Page 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Page 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Page 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...