Potential cause
Computer freeze-up or connection failure.
Recommended actions
1. Attempt a reboot.
2. System reboots successfully.
3. Contact (
Potential cause
PC or HMI failure.
Recommended actions
1. Attempt a reboot.
2. System does not reboot.
3. Contact Cascade (
).
7.5.7
Other error not covered in other sections
Potential cause
Unknown
Recommended actions
Contact Cascade (
7.6
Replacing the power socket fuses
The analyzer contains two external fuses located in the power socket attached to the rear
panel.
DANGER
ELECTRIC SHOCK
The analyzer operates using mains voltage that is dangerous to life. Make sure that the
power ON/OFF switch at the rear of the panel is set to OFF and the power cable plug is
removed.
The analyzer MUST be earthed using a grounded three core cable suitable for the rating
defined in the General Characteristics (
Death or personal injury may result if this is not observed.
Procedure
1. Turn the
ON/OFF
switch on the back panel to
OFF
.
2. Disconnect the external power supply from the power socket.
3. Allow the analyzer to cool.
Reference Manual
Troubleshooting and diagnostics
00809-0100-4244
January 2020
Reference Manual
83
Summary of Contents for Rosemount CT4400
Page 1: ...Reference Manual 00809 0100 4244 Rev AA January 2020 Rosemount CT4400 Continuous Gas Analyzer ...
Page 32: ...Start up procedure Reference Manual January 2020 00809 0100 4244 32 Rosemount CT4400 ...
Page 40: ...Operating the analyzer Reference Manual January 2020 00809 0100 4244 40 Rosemount CT4400 ...
Page 64: ...Gas calibration Reference Manual January 2020 00809 0100 4244 64 Rosemount CT4400 ...
Page 96: ...Shutdown procedure Reference Manual January 2020 00809 0100 4244 96 Rosemount CT4400 ...
Page 98: ...Preventative maintenance Reference Manual January 2020 00809 0100 4244 98 Rosemount CT4400 ...
Page 100: ...Recommended spare parts Reference Manual January 2020 00809 0100 4244 100 Rosemount CT4400 ...
Page 105: ......