Standardize Calibration: Standardizing the sensor to a known turbidity.
•
Grab Calibration: Standardizing the sensor to a known turbidity based on a reference
turbidmeter.
7.2.1
Calibrate slope
Conduct a two-point calibration of the turbidity sensor against a user-prepared
nephelometric turbidity unit (NTU) standard.
Procedure
1. Immerse the sensor in filtered water having very low turbidity and measure the
sensor output.
2. Increase the turbidity of the filtered water by a known amount, typically 20 NTU,
and measure the sensor output again.
The transmitter takes the two measurements, applies a linear correction (if
necessary), and calculates the sensitivity.
Sensitivity is the sensor output (in mV) divided by turbidity. A typical new sensor has a
turbidity of about 10 mV/NTU. As the sensor ages, the sensitivity decreases. Dark current
is the signal generated by the detector when no light is falling on it. The transmitter
substracts the dark current from the raw scattered light signal and converts the result to
turbidity. In highly filtered samples, which scatter little light, the dark current can be a
substantial amount of the signal generated by the detector.
7.2.2
Calibrate against a standard
You can calibrate the turbidity sensor against a commercial standard.
Stable 20.0 nephelometric turbidity unit (NTU) standards are available from a number of
sources. Calibration using a commercial standard is simple. Filtered deionized water is not
required.
Prerequisites
Before beginning the calibration, the transmitter does a dark current measurement. Dark
current is the signal generated by the detector even when no light is falling on it.
Procedure
1. Select
Standard Calibration
.
2. Press
ENTER/MENU
.
The transmitter subtracts the dark current from the raw scattered light signal and
converts the result to turbidity. In highly filtered samples, which scatter little light,
the dark current can be a substantial amount of the signal generated by the sensor.
Calibrate
Reference Manual
July 2020
00809-0100-3565
46
Emerson.com/Rosemount
Summary of Contents for Rosemount Clarity II T56
Page 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Page 4: ...4 ...
Page 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Page 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Page 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Page 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Page 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...