33
Reference Manual
00809-0300-4750, Rev CA
Quick Installation and Start-Up
March 2016
Quick Installation and Start-Up
Flow units (PV)
The
flow units
variable specifies the format in which the flow rate will be displayed. Units
should be selected to meet your particular metering needs. See
for available
units of measure.
Line size
The
line size
(sensor size) must be set to match the actual sensor connected to the
transmitter. The size must be specified in inches. See
for available sensor sizes.
URV (Upper Range Value)
The
upper range value
(URV) sets the 20 mA point for the analog output. This value is
typically set to full-scale flow. The units that appear will be the same as those selected under
the
flow units
parameter. The URV may be set between –39.3 ft/s to 39.3 ft/s
(–12 m/s to 12 m/s). There must be at least 1 ft/s (0.3 m/s) span between the URV and LRV.
LRV (Lower Range Value)
The
lower range value
(LRV) sets the 4 mA point for the analog output. This value is typically
set to zero flow. The units that appear will be the same as those selected under the
flow units
parameter. The LRV may be set between -39.3 ft/s to 39.3 ft/s (-12 m/s to 12 m/s). There
must be at least 1 ft/s (0.3 m/s) span between the URV and LRV.
LOI menu path
Field Mount: Basic Setup, Flow Units, PV Units
Wall Mount: UNITS
Traditional Fast Keys
1, 3, 1
Device dashboard
2, 2, 1, 2
LOI menu path
Field Mount: Basic Setup, Line Size
Wall Mount: TUBE SIZE
Traditional Fast Keys
1, 3, 1
Device dashboard
2, 2, 1, 4, 2
LOI menu path
Field Mount: Basic Setup, PV URV
Wall Mount: ANALOG OUTPUT RANGE
Traditional Fast Keys
1, 3, 1
Device dashboard
2, 2, 1, 3, 3
LOI menu path
Field Mount: Basic Setup, PV LRV
Wall Mount: AUX. FUNCTION
Traditional Fast Keys
1, 3, 1
Device dashboard
2, 2, 1, 3, 2
Summary of Contents for Rosemount 8750W
Page 2: ......
Page 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Page 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Page 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Page 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Page 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Page 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Page 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Page 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Page 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Page 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Page 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Page 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Page 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Page 263: ......