123
Reference Manual
00809-0300-4750, Rev CA
Advanced Diagnostics Configuration
March 2016
Advanced Diagnostics Configuration
6.8
4-20 mA loop verification
The
4-20 mA loop verification
diagnostic provides a means of verifying the analog output
loop is functioning properly. This is a manually initiated diagnostic test. This diagnostic
checks the integrity of the analog loop and provides a health status of the circuit. If the
verification does not pass, this will be highlighted in the results given at the end of the
check. This diagnostic is only available on the Field Mount Transmitter.
The
4-20 mA loop verification
diagnostic is useful for testing the analog output when errors
are suspected. The diagnostic tests the analog loop at five different mA output levels:
4 mA
12 mA
20 mA
Low alarm level
High alarm level
6.8.1
Initiating 4-20 mA loop verification
The
4-20 mA loop verification
diagnostic can be initiated as required by the application. If the
advanced diagnostics suite 2 (DA2 Option) was ordered, then the
4-20 mA loop verification
diagnostic will be available. If DA2 was not ordered or licensed, this diagnostic is not
available.
6.8.2
4-20 mA loop verification parameters
The
4-20 mA loop verification
diagnostic has five read-only parameters plus an overall test
result. It does not have any configurable parameters.
4-20 mA loop verification test result
Shows the results of the
4-20 mA loop verification
test as either passed or failed.
LOI menu path
Field Mount: Diagnostics, Advanced Diag, 4-20mA Verify, 4-20mA Verify
Wall Mount: N/A
Traditional Fast Keys
1, 2, 3, 3, 1
Device dashboard
3, 4, 3, 1
LOI menu path
Field Mount: Diagnostics, Advanced Diag, 4-20mA Verify, View Results
Wall Mount: N/A
Traditional Fast Keys
1, 2, 3, 3, 2
Device dashboard
3, 4, 3
Summary of Contents for Rosemount 8750W
Page 2: ......
Page 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Page 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Page 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Page 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Page 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Page 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Page 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Page 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Page 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Page 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Page 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Page 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Page 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Page 263: ......