Reference Manual
00809-0300-4750, Rev CA
Advanced Configuration Functionality
March 2016
Advanced Configuration Functionality
The
time limit
parameter forces the output and running average values to the new value of
an actual flow rate change that is outside the percent limit boundaries. It thereby limits
response time to flow changes to the time limit value rather than the length of the running
average.
For example, if the number of samples selected is 100, then the response time of the
system is 10 seconds. In some cases this may be unacceptable. By setting the time limit, you
can force the Rosemount 8750W to clear the value of the running average and re-establish
the output and average at the new flow rate once the time limit has elapsed. This parameter
limits the response time added to the loop. A suggested time limit value of two seconds is a
good starting point for most applicable process fluids. This parameter can be configured
between zero and 256 seconds. The default value is two seconds.
5.5
Configure special units
Special units are used when the application requires units that are not included in the flow
units available from the device. Refer to
for a complete list of the
available units.
5.5.1
Base volume unit
Base volume unit
is the unit from which the conversion is being made. Set this variable to the
appropriate option.
5.5.2
Conversion factor
The special units
conversion factor
is used to convert base units to special units. For a straight
conversion of units from one unit of measure to a different unit of measure, the
conversion
factor
is the number of base units in the new unit.
For example, if you are converting from gallons to barrels and there are 31 gallons in a
barrel, the conversion factor is 31.
LOI menu path
Field Mount: Basic Setup, Flow Units, Special Units, Base Vol Units
Wall Mount: AUX. FUNCTION, Special Units, Base Vol Units
Traditional Fast Keys
1, 3, 2, 2,2
Device dashboard
2, 2, 1, 6
LOI menu path
Field Mount: Basic Setup, Flow Units, Special Units, Conv Factor
Wall Mount: AUX. FUNCTION, Special Units, Conv Factor
Traditional Fast Keys
1, 3, 2, 2, 3
Device dashboard
2, 2, 1, 6
Summary of Contents for Rosemount 8750W
Page 2: ......
Page 96: ...84 Reference Manual 00809 0300 4750 Rev CA Operation March 2016 Operation ...
Page 168: ...156 Reference Manual 00809 0300 4750 Rev CA Maintenance March 2016 Maintenance ...
Page 214: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 202 Wiring Diagrams ...
Page 218: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 206 Wiring Diagrams ...
Page 219: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 207 Wiring Diagrams ...
Page 220: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 208 Wiring Diagrams ...
Page 222: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 210 Wiring Diagrams ...
Page 223: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 211 Wiring Diagrams ...
Page 224: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 212 Wiring Diagrams ...
Page 225: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 213 Wiring Diagrams ...
Page 227: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 215 Wiring Diagrams ...
Page 228: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 216 Wiring Diagrams ...
Page 229: ...Wiring Diagrams March 2016 Reference Manual 00809 0300 4750 Rev CA 217 Wiring Diagrams ...
Page 263: ......