56
Reference Manual
00809-0100-4697, Rev GA
Section 4: Operation and Maintenance
November 2014
Operation and Maintenance
4.4.2
Resource block
4.4.3
Transducer block troubleshooting
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
Memory Failure
BLOCK_ERR will show the lost NV Data or Lost Static Data bit set.
Restart the device by setting RESTART to Processor. If the block
error does not clear, call the factory.
Block Alarms Will not
work
Features
FEATURES_SEL does not have Alerts enabled. Enable the report bit.
Notification
LIM_NOTIFY is not high enough. Set equal to MAX_NOTIFY.
Symptom
Possible causes
Corrective action
Mode will not leave
OOS
Target mode not set
Set target mode to something other than OOS.
A/D board check sum error
The A/D board has a checksum error.
Resource block
The actual mode of the Resource block is in OOS. See Resource
Block Diagnostics for corrective action.
Transducer Block
The actual mode of the Transducer Block is OOS.
The primary value is
BAD
Measurement
Look at the SENSOR_STATUS parameter (See
)