16. Verify that the level gauge head is properly aligned. The cover on the terminal
compartment shall be parallel to the Verification Pin. The notch on the still-pipe
flange indicates the direction of the Verification Pin.
%
$
&
'
(
)
*
A. Rosemount 5900 Level Gauge
B. Pressure transmitter
C. Mark that indicates direction of the Verification Pin
D. Cover on the terminal compartment
E. Parallel
F. Verification Pin
G. Still-pipe
17. Wire the gauge, and configure by using the Rosemount TankMaster WinSetup
software as described in the Rosemount Tank Gauging
18. Configure the level gauge for LPG measurements (see
).
Installation
Reference Manual
September 2019
00809-0100-5900
76