3. Using the box provided on the nameplate, the User shall permanently mark the
type of protection chosen for the specific installation. Once the type of protection
has been marked it shall not be changed.
4. When installed as Ex ib Ga/Gb, the partition wall materials separating EPL Ga from
EPL Gb are constructed of different materials depending on the antenna option.
Please refer to Control Drawing D9240040-917 for the material type of each
antenna. The material shall not be subject to environmental conditions which might
adversely affect the partition wall.
5. Maximum Process Temperatures are as follows:
When option n=Tank Seal
O-ring Type
Min/Max Process
Temperature Range
PV or QV
Viton
-15°C to +180°C
PK, FK, HK or QK
Kalrez
-20°C to +230°C
PE or QE
EPDM
-40°C to +110°C
PB or QB
BUNA-N
-35°C to +90°C
PM, FF, HH or QM
FVMQ
-60°C to +155°C
PF or QF
FEP
-60°C to +180°C
B.7
Europe
B.7.1
I1 ATEX Intrinsic Safety
Certificate
FM09ATEX0057X
Standards
EN IEC 60079-0:2018, EN 60079-11:2012, EN 60079-26:2015, EN
60529:1991+A1(2000)+A2(2013)
Markings
II 1 G Ex ia IIC T4 Ga
II 1/2 G Ex ib IIC T4 Ga/Gb
Ta = -50°C to 80°C; IP66, IP67
Ui (Vmax)
Ii (lmax)
Pi
Ci
Li
Entity parameters
30 V
300 mA
1.3 W
1.1 nF
1.5 µH
FISCO parameters
17.5V
380 mA
5.32 W
1.1 nF
1.5 µH
Specific Conditions for Safe Use (X):
1. The enclosure contains aluminum and is considered to present a potential risk of
ignition by impact or friction. When installed as EPL Ga, care must be taken during
installation and use to prevent impact or friction.
Product Certifications
Reference Manual
September 2019
00809-0100-5900
256
Reference Manual
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...