A.1.8
Installation considerations
.
A.1.9
Measurement principle
The FMCW-method (Frequency Modulated Continuous Wave) means that the transmitted
radar signal has a linear frequency variation around 10 GHz. The reflection from the liquid
surface has a slightly different frequency compared with the signal transmitted from the
antenna when the reflection is received. The difference in frequency is directly
proportional to the distance between the antenna and the liquid surface, and thereby also
the liquid level. This technology enables a very accurate and stable measured value.
Figure A-1: Principle of FMCW Technology
$
%
A. Frequency, f (GHz)
B. Time, t (s)
A.2
Communication / Display / Configuration
A.2.1
Output variables and units
•
Level, and ullage: meter, centimeter, millimeter, feet, or inch
•
Level rate: meter/second, meter/hour, feet/second, feet/hour, inch/minute
•
Signal strength: mV
A.2.2
Configuration tools
Rosemount TankMaster WinSetup, Field Communicator
A.3
F
OUNDATION
™
Fieldbus characteristics
Polarity sensitive
No
Quiescent current draw
51 mA
Specifications and reference data
Reference Manual
September 2019
00809-0100-5900
224
Reference Manual
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...