Choose correction method
There are several options available depending on gas mixture in the tank.
Procedure
1. In the
LPG Setup
window click the
Correction
button to open the
LPG Correction
window:
2. Choose one of the following correction methods:
Option
Description
Air Correction, LPG
correction disabled
This method should only be used when there is no vapor in
the tank, i.e. when the tank is empty and contains air only.
It is used in the initial step when calibrating the Rosemount
5900S.
One known gas
This method may be used when there is only one gas type
in the tank. It provides the highest accuracy among the
different correction methods. Note that even small
amounts of another gas reduces the accuracy.
One or more
unknown gases
Use this method for hydrocarbons, for example Propane/
Buthane, when the exact mixture is not known.
Two gases with
unknown mixratio
This method is suitable for a mixture of two gases even if
the mixratio is not known.
One or more known
gases with known
mixratio
This method may be used when there is a well known
mixture of up to 4 products in the tank.
Now the Rosemount 5900S level gauge is ready to measure the product level when
the tank is put into operation.
Configuration
Reference Manual
September 2019
00809-0100-5900
128
Reference Manual
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...