Reference Manual

00809-0100-3550, Rev AA

March 2019

Rosemount

 550pH

Single-use pH Sensor for Bioprocessing Applications

Summary of Contents for Rosemount 550pH

Page 1: ...Reference Manual 00809 0100 3550 Rev AA March 2019 Rosemount 550pH Single use pH Sensor for Bioprocessing Applications ...

Page 2: ...rence If you do not understand any of the instructions contact your Emerson representative for clarification Follow all warnings cautions and instructions marked on and supplied with the product Inform and educate your personnel in the proper installation operation and maintenance of the product Install equipment as specified in the installation instructions of the appropriate Reference Manual and...

Page 3: ...sor 9 Chapter 3 Wire the sensor 13 Chapter 4 Calibrate 17 4 1 One point standardization based on a grab sample 17 Chapter 5 Troubleshoot 19 5 1 Faulty reading 19 5 2 Label information too small 19 Chapter 6 China RoHS table 21 Chapter 7 EU Declaration of Conformity 23 Reference Manual Contents 00809 0100 3550 March 2019 Reference Manual 3 ...

Page 4: ...Contents Reference Manual March 2019 00809 0100 3550 4 Rosemount 550pH ...

Page 5: ...units Operating temperature range 39 to 104 F 4 to 40 C Pressure range 0 to 30 Psig 0 to 2 bar pH range 2 to 12 Sensor drift Less than 0 005 pH drift per day in laboratory conditions Initial one point standardization accuracy using internal sensor buffer Less than 0 1 pH at pH 7 Shelf life 2 years at 77 F 25 C expiration date on sensor label Measuring method Potentiometry pH glass electrode and do...

Page 6: ...units Electrical connector Variopol VP 8 0 Sensitivity Greater than 57 mV pH at 77 F 25 C Electrolyte Phosphate based see SDS on product website Integral temperature sensor Pt 100 Gamma irradiation Up to 50 kGy Sensor readiness Reference Manual March 2019 00809 0100 3550 6 Rosemount 550pH ...

Page 7: ...onal drawings See Figure 1 1 for sensor dimensions Figure 1 1 Rosemount 550pH A1M22E0VP000 Dimensional Drawing A VP8 cable connection Reference Manual Sensor readiness 00809 0100 3550 March 2019 Reference Manual 7 ...

Page 8: ...Sensor readiness Reference Manual March 2019 00809 0100 3550 8 Rosemount 550pH ...

Page 9: ... to Rosemount transmitters 3 Use the Calibration menu on the transmitter to go to Slope offset setting Recommendation Do not fill the bag with buffer or other solution until after 3 b to confirm sensor functionality a Input the factory calibrated slope on the sensor label into the Slope at 25 C field and press ENTER Press EXIT to go back to the Calibration menu Do not touch the value in the Offset...

Page 10: ... to complete the initial sensor standardization Note Use pH sensor prior to expiration date See Figure 2 2 4 Clip the zip tie locking the clip in place Remove the circular warning tag if it is still present See Figure 2 3 Figure 2 3 Clipped Zip Tie 5 Press the slide lock into the unlocked position The small lip of the lock should be the side that is pressed in See Figure 2 4 Install the sensor Ref...

Page 11: ...on 6 Use the plunger assembly to push the sensor into the activated position See Figure 2 5 Figure 2 5 Activated Sensor 7 Press the slide lock into the locked position The small tab on the slide lock should be protruding out See Figure 2 6 Reference Manual Install the sensor 00809 0100 3550 March 2019 Reference Manual 11 ...

Page 12: ...gure 2 7 Zip Tie Postrequisites After the start up is complete there is no additional work needed on the sensor unless a one point standardization is deemed necessary See One point standardization based on a grab sample for this procedure using an off line sample Install the sensor Reference Manual March 2019 00809 0100 3550 12 Rosemount 550pH ...

Page 13: ...See Figure 3 1 to wire the Rosemount 550pH Sensor to the Rosemount 56 1056 or 1057 Transmitter Figure 3 1 Wire the Rosemount 550pH Sensor to the Rosemount 56 1056 or 1057 Transmitter 2 See Figure 3 2 to wire the Rosemount 550pH Sensor to the Rosemount 1066 Transmitter Reference Manual Wire the sensor 00809 0100 3550 March 2019 Reference Manual 13 ...

Page 14: ...Rosemount 550pH Sensor to the Rosemount 1066 Transmitter 3 See Figure 3 3 to wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter Wire the sensor Reference Manual March 2019 00809 0100 3550 14 Rosemount 550pH ...

Page 15: ...Figure 3 3 Wire the Rosemount 550pH Sensor to the Rosemount 5081 Transmitter Reference Manual Wire the sensor 00809 0100 3550 March 2019 Reference Manual 15 ...

Page 16: ...Wire the sensor Reference Manual March 2019 00809 0100 3550 16 Rosemount 550pH ...

Page 17: ...ndardize grab in the Calibration menu 4 Once the pH reading of the sensor is stabilized press ENTER to pass the installation prompt 5 Input the pH of the grab sample into the pH of grab sample field and press ENTER 6 Press EXIT to complete the calibration The pH reading on the screen should now equal the pH of the grab sample Note pH is a temperature dependent attribute Please ensure off line samp...

Page 18: ...Calibrate Reference Manual March 2019 00809 0100 3550 18 Rosemount 550pH ...

Page 19: ...ithin sensor sleeve 90 degrees 2 Repeat sensor standardization procedure 5 2 Label information too small The serial number Cal pH or slope are too small to read Recommended action Scan the two dimensional bar code on the sensor label to retrieve information Reference Manual Troubleshoot 00809 0100 3550 March 2019 Reference Manual 19 ...

Page 20: ...Troubleshoot Reference Manual March 2019 00809 0100 3550 20 Rosemount 550pH ...

Page 21: ...6 China RoHS table Reference Manual China RoHS table 00809 0100 3550 March 2019 Reference Manual 21 ...

Page 22: ...China RoHS table Reference Manual March 2019 00809 0100 3550 22 Rosemount 550pH ...

Page 23: ...7 EU Declaration of Conformity Reference Manual EU Declaration of Conformity 00809 0100 3550 March 2019 Reference Manual 23 ...

Page 24: ...EU Declaration of Conformity Reference Manual March 2019 00809 0100 3550 24 Rosemount 550pH ...

Page 25: ...Reference Manual 00809 0100 3550 March 2019 Reference Manual 25 ...

Page 26: ...ST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai United Arab Emirates P O Box 17033 T 971 4 811 8100 F 971 4 886 5465 liquid csc emerson com ASIA PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore T 65 777 8211 F 65 777 0947 liquid csc emerson com Linkedin com company Emerson Automation Solutions twitter com rosemount_news Facebook com...