7.9.9
Open the configuration report
The configuration report shows what configuration changes have been done to the
transmitter compared to the factory configuration. The report compares a specified
backup file with the default transmitter configuration.
Procedure
1. In Rosemount Radar Master, select
Tools
→
Configuration Report
.
2. Browse to desired backup file to generate the configuration report from, and then
select
Open
.
Information on probe type, software versions, software and hardware
configuration, and unit code is presented.
Figure 7-24: The Configuration Report Window in Rosemount Radar Master
7.9.10
Reset to factory settings
Reset to factory settings using Rosemount Radar Master
This function resets all or a specific part of the holding registers to factory settings.
Prerequisites
Note
It is recommended that a backup of the configuration is done before the factory reset.
Then the old transmitter configuration can be loaded if necessary.
Procedure
Select
Tools
→
Factory Settings
.
Service and troubleshooting
Reference Manual
April 2019
00809-0100-4530
216
Rosemount 5300 Level Transmitter