Quick Start Guide

00825-0200-3499, Rev AB

November 2019

Rosemount

 499ACL-02

Total Chlorine Sensor

Summary of Contents for Rosemount 499ACL-02

Page 1: ...Quick Start Guide 00825 0200 3499 Rev AB November 2019 Rosemount 499ACL 02 Total Chlorine Sensor ...

Page 2: ...ING Physical access Unauthorized personnel may potentially cause significant damage to and or misconfiguration of end users equipment This could be intentional or unintentional and needs to be protected against Physical security is an important part of any security program and fundamental to protecting your system Restrict physical access by unauthorized personnel to protect end users assets This ...

Page 3: ...Figure 1 1 Rosemount 499ACL 02 Sensor Parts A Membrane retainer B Membrane assembly C O ring D Cathode E Electrolyte fill plug wrap with pipe tape F Pressure equalizing port G Sensor cable integral cable shown 1 3 Specifications Table 1 1 Sensor Specifications Physical characteristics Specifications Pressure 0 to 65 psig 101 to 549 kPa abs Temperature operating 32 to 122 F 0 to 50 C Process connec...

Page 4: ...Table 1 1 Sensor Specifications continued Physical characteristics Specifications Cathode Gold Linearity 2 typical Quick Start Guide November 2019 4 Emerson com Rosemount ...

Page 5: ... for use only in the Rosemount TCL sample conditioning system for total chlorine Consult the instruction manual for the Rosemount TCL sample conditioning system for more information November 2019 Quick Start Guide Quick Start Guide 5 ...

Page 6: ...t 1056 and 56 Transmitters Table 3 1 Rosemount 499ACL 02 Sensor Wiring to Rosemount 1056 and 56 Transmitters Terminal number Letter Wire color Description 1 A White Resistance temperature device RTD return 2 B White red RTD sense 3 C Red RTD in 4 D Clear RTD shield 5 N A N A 5 V out 6 N A N A 4 5 V out 7 D Clear Anode shield 8 E Gray Anode 9 D Clear Cathode shield Quick Start Guide November 2019 6...

Page 7: ...escription 10 F Orange Cathode Figure 3 2 Rosemount 499ACL 02 Sensor Wiring to Rosemount 5081 Transmitter Table 3 2 Rosemount 499ACL 02 Sensor Wiring to Rosemount 5081 Transmitter Terminal number Letter Wire color Description 1 N A N A N A 2 N A N A 0 8 V 3 A White RTD return November 2019 Quick Start Guide Quick Start Guide 7 ...

Page 8: ...ite red RTD sense 5 C Red RTD in 6 N A N A Reference guard 7 N A N A Reference in 8 D Clear Solution ground 9 N A N A pH guard 10 N A N A pH in 11 N A N A 5 V 12 N A N A 5 V 13 E Gray Anode 14 F Orange Cathode 15 N A N A HART FOUNDATION Fieldbus 16 N A N A HART FOUNDATION Fieldbus Quick Start Guide November 2019 8 Emerson com Rosemount ...

Page 9: ...es to solution ground terminal on TB 2 Use wire nut and pigtail if necessary Table 3 3 Rosemount 499ACL 02 Wiring to Rosemount 1066 Transmitter Letter Color Terminal description A Orange Cathode B Gray Anode C White Return D White red Sense E Red RTD in November 2019 Quick Start Guide Quick Start Guide 9 ...

Page 10: ...gram Terminal number Description 1 Cathode 2 N A 3 RTD sense 4 Anode 5 RTD return 6 RTD in When making a connection through a junction box PN 23550 00 wire point to point NOTICE Use a wire nut and pigtail included when connecting several wires to the same terminal Quick Start Guide November 2019 10 Emerson com Rosemount ...

Page 11: ...4 Follow the transmitter prompts for zeroing the sensor The zero current should be between 10 and 50 nA For more information refer to the transmitter manual 4 2 Full scale calibration Because stable dilute chlorine standards are not available the sensor must be calibrated against the results of a laboratory test run on a grab sample of the process liquid Procedure 1 Place the sensor in the flow ce...

Page 12: ... pressure is reduced to 0 psig and the process temperature is lowered to a safe level 5 1 Cleaning the membrane Keep the membrane clean Clean the membrane with water sprayed from a wash bottle Use a soft tissue to gently wipe the membrane 5 2 Replacing the electrolyte solution and membrane WARNING Harmful substance Fill solution may cause irritation May be harmful if swallowed Read and follow the ...

Page 13: ...at the base of the cathode stem with several drops of electrolyte solution 17 Insert a small blunt probe like a toothpick with the end cut off through the pressure equalizing port See Figure 1 1 CAUTION Equipment damage A sharp probe may puncture the bladder and destroy the sensor Do not use a sharp probe 18 Gently press the probe against the bladder several times to force liquid through the holes...

Page 14: ...rane retainer 23501 02 Total chlorine membrane assembly includes one membrane assembly and one O ring 23502 02 Total chlorine membrane kit includes three membrane assemblies and three O rings 9210438 Total chlorine sensor fill solution 4 oz 120 mL Quick Start Guide November 2019 14 Emerson com Rosemount ...

Page 15: ...November 2019 Quick Start Guide Quick Start Guide 15 ...

Page 16: ...LE EAST AND AFRICA Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai United Arab Emirates P O Box 17033 971 4 811 8100 971 4 886 5465 liquid csc emerson com ASIA PACIFIC Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore 65 6 777 8211 65 6 777 0947 liquid csc emerson com Linkedin com company Emerson Automation Solutions twitter com rosemount_news...