Figure 3-57: CPU or Optional I/O Module replacement
A. Terminal end of Transmitter Electronics Enclosure
B. CPU Module
C. Optional I/O Module
D. Power Supply board
E. Fuse
F. Internal chassis ground
5. Grasp the module you want to replace and pull it out of the enclosure.
6. Insert the new CPU Module or I/O Module into the enclosure and firmly push until
the board is fully seated into the Backplane Board connectors.
7. Replace the terminal blocks for the CPU Module and/or the Optional I/O Module
and tighten the flat head screws with a 3mm flat blade screw driver.
8. Connect to the meter with MeterLink. Open the Meter Monitor (Detailed) view and
verify the meter is acquiring data, the transducers have good signals and flow
profiles for the chords are displayed and run the Zero Flow Calibration Wizard;
select the
Calibrate
→
Zero Calibrate menu
and generate the report. Then, collect
and save a Maintenance log to show the “as left” condition of the meter. Or use
AMS
™
Device Manager or a HART
®
Field Communicator and select
Overview
→
Zero Calibration
→
Start menu
or
Service Tools
→
Maintenance
→
Zero
Calibration
→
Start
.
9. Close your application connection (MeterLink, AMS
™
Device Manager or HART
®
Field Communicator).
10. If you are not replacing other electronics, replace the end caps and security latches
(requires a 3 mm Allen wrench).