Table 6-4: Maintenance Alerts (M:)
(continued)
Message
Description
Recommended actions
Non-PV Out of Limits
One of the non-primary
measurements is outside the
associated sensor limits and may be
unreliable.
1. Bring the system to a safe state.
2. Verify that all non-primary
measurements are within
specified limits.
3. Restart the device.
4. If the condition persists, replace
the device
Table 6-5: Advisory Alerts (A:)
Message
Description
Recommended actions
Database Memory Warning
The device has failed to write to the
database memory at some time in the
past. Any data written during this time
may have been lost.
1. If logging dynamic data is not
needed, this advisory alert can be
safely ignored.
2. Restart the device.
3. Reconfirm all configuration items
in the device.
4. Restore default settings and
reconfigure the device.
5. If the condition persists, replace
the device.
Non-Critical User Data Warning
A user written parameter does not
match expected value.
1. Restart the device.
2. Reconfirm all configuration items
in the device.
3. Restore default settings and
reconfigure the device.
4. If the condition persists, replace
the device.
Volume Range Warning
The level measurement is outside the
configured volume range.
•
Check volume configuration.
Verification Mode Active
The device is in verification mode and
is not reporting actual information.
•
If this behavior is not desired, stop
verification mode.
Button Stuck
The button on the Electronics Board is
detected as stuck in the active
position.
1. Check the buttons for
obstructions.
2. If the condition persists, restart
the device.
3. If the condition persists, replace
the device.
Reference Manual
Service and troubleshooting
00809-0100-4308
February 2020
Reference Manual
113
Summary of Contents for Rosemount 3308A
Page 8: ...Contents Reference Manual February 2020 00809 0100 4308 viii Rosemount 3308 Series ...
Page 10: ...Introduction Reference Manual February 2020 00809 0100 4308 10 Rosemount 3308 Series ...
Page 22: ...Transmitter overview Reference Manual February 2020 00809 0100 4308 22 Rosemount 3308 Series ...
Page 74: ...Installation Reference Manual February 2020 00809 0100 4308 74 Rosemount 3308 Series ...