![Emerson Rosemount 3152 Reference Manual Download Page 80](http://html.mh-extra.com/html/emerson/rosemount-3152/rosemount-3152_reference-manual_104708080.webp)
Rosemount 3150 Series
Reference Manual
00809-0100-4835 Rev BE
February 2018
71
Transmitter Spare Parts
GENERAL
CONSIDERATIONS
Because of the nuclear use intended for these parts, certain
factors must be considered regarding maintenance of product
qualification and component traceability during on-site
instrument repair. Rosemount Nuclear rigidly controls the
manufacture of each instrument to ensure that published
performance specifications are met and qualified
configurations are maintained. For parts installed outside of
this controlled environment, Rosemount Nuclear is unable to
ensure that the specifications are being satisfied. This
responsibility is shifted to the end user. The integrity of the
instrument as originally assembled is modified.
Replacement of parts has ramifications under 10CFR21, for
which the user is responsible. These same regulations also
mandate a component traceability program, which the user
must undertake for the replacement parts. In view of this, and
to maintain the qualification of the product, the user must
ensure that all replacement parts are installed in accordance
with the Rosemount Nuclear approved installation and
calibration procedures herein.
NOTE
Spare parts for Rosemount 3152, 3153 and 3154 models are
not hydrostatically tested or nuclear cleaned.
NOTE
The part numbers shown are current at the time of printing of
this manual, but may be revised in the future. Parts provided
are compatible and interchangeable with those listed on your
order as to the form, fit, and function of the part required.
SPARE PARTS STORAGE
AND SHELF LIFE
Store all spare transmitters and spare component parts in
accordance with ANSI N45.2.2 level B.
Qualified transmitters, spare electronic assemblies, and spare
terminal blocks were qualified based on a shelf life of 20 years
at an ambient temperature of 90°F (32.2°C).
NOTE
Spare electronic assemblies should be stored in the original
packaging as shipped by Rosemount Nuclear.
Spare O-rings: Shelf life is 40 years at an ambient
temperature of 90°F (32.2°C).
Lubricants and sealants: The date of the end of shelf life (use
by date) is provided with the lubricants and/or sealants, at the
time of shipment. The product has a minimum of six months
shelf life at the time of shipment.
All other parts: Shelf life is not applicable.