171
Reference Manual
00809-0100-4021, Rev JA
Operation and Maintenance
April 2017
Operation and Maintenance
Replacing the electronics module
Use the following procedure to reassemble the electronics housing for the transmitter:
1. Examine the electronics module to ensure that the failure mode and transmitter security switches are
in the desired positions.
2. Carefully insert the electronics module lining up the interconnecting pins with the necessary
receptacles on the electronics board.
3. Tighten the two mounting screws. Replace the LCD display, if applicable.
4. Replace the cover. Tighten
1
/
6
of a revolution after the cover begins to compress the O-ring. Both
transmitter covers must be fully engaged to meet explosion proof requirements.
5.2.4
Transmitter diagnostics logging
The Transmitter Diagnostics Logging feature stores advanced diagnostic information between device
resets, such as what caused the transmitter to go into alarm, even if that event has disappeared. For
example, if the transmitter detects an open sensor from a loose terminal connection, the transmitter will
go into alarm. If wire vibration causes that wire to begin making a good connection, the transmitter will
come out of alarm. This jumping in and out of alarm is frustrating when trying to determine what is
causing the problem. However, the Transmitter Diagnostics Logging feature keeps track of what caused
the transmitter to go into alarm and saves valuable debugging time. The log may be viewed using an
asset management software, such as AMS Device Manager.
5.3
Return of materials
To expedite the return process in North America, call the Emerson National Response Center
(800-654-7768) for assistance with any needed information or materials.
The center will ask for the following information:
Product model
Serial numbers
The last process material to which the product was exposed
The center will provide
A Return Material Authorization (RMA) number
Instructions and procedures to return goods that were exposed to hazardous substances
For other locations, contact an Emerson representative.
Note
If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be
available to people exposed to specific hazardous substances, must be included with the returned
materials.
Summary of Contents for Rosemount 3144P
Page 2: ......
Page 10: ...8 Reference Manual 00809 0100 4021 Rev JA Title Page April 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 4021 Rev JA Introduction April 2017 Introduction ...
Page 38: ...28 Reference Manual 00809 0100 4021 Rev JA Installation April 2017 Installation ...
Page 235: ...225 Reference Manual 00809 0100 4021 Rev JA April 2017 ...