Reference Manual
00809-0100-4801, Rev FA
October 2010
Rosemount 3051S Series
2-18
Electrical Considerations
Proper electrical installation is necessary to prevent errors due to improper
grounding and electrical noise. For Junction Box housing, shielded signal
wiring should be used in high EMI/RFI environments.
Remote Display Wiring
and Power Up
The Remote Mount Display and Interface system consists of a local
transmitter and a remote mount LCD display assembly. The local 3051S
transmitter assembly includes a Junction Box housing with a three position
terminal block integrally mounted to a SuperModule. The remote mount LCD
display assembly consists of a dual compartment PlantWeb housing with a
seven position terminal block. See Figure 2-10 on page 2-19 for complete
wiring instructions. The following is a list of necessary information specific to
the Remote Mount Display system:
• Each terminal block is unique for the remote display system.
• A 316 SST housing adapter is permanently secured to the remote
mount LCD display PlantWeb housing providing an external ground
and a means for field mounting with the provided mounting bracket.
• A cable is required for wiring between the transmitter and remote
mount LCD display. The cable length is limited to 100-ft.
• 50-ft. (option M8) or 100-ft. (option M9) cable is provided for wiring
between the transmitter and remote mount LCD display. Option M7
does not include cable; see recommended specifications below:
Cable type:
Recommend Madison AWM Style 2549 cable. Other comparable
cable may be used as long as it has independent dual twisted shielded pair
wires with an outer shield. The Power wires must be 22 AWG minimum and
the CAN communication wires must be 24 AWG minimum.
Cable length:
Up to 100-ft. (31 m) depending upon cable capacitance.
Cable capacitance:
The capacitance as wired must be less than 5000
picofarads total. This allows up to 50 picofarads per ft. (0.3 m) for a 100-ft. (31
m) cable.
Intrinsic Safety Consideration:
The transmitter assembly with remote
display has been approved with Madison AWM Style 2549 cable. Alternate
cable may be used as long as the transmitter with remote display and cable is
configured according to the installation control drawing or certificate. Refer to
appropriate approval certificate or control drawing in Appendix B for remote
cable IS requirements.
IMPORTANT
Do not apply power to the remote communications terminal. Follow wiring
instructions carefully to prevent damage to system components.
IMPORTANT
For ambient temperatures above 140 °F (60 °C), cable wiring must be rated at
least 9 °F (5 °C) above the maximum ambient temperature.
Summary of Contents for Rosemount 3051S Series
Page 2: ......
Page 4: ......
Page 10: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series TOC 6 ...
Page 14: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series 1 4 ...
Page 74: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series 3 32 ...
Page 90: ...Reference Manual 00809 0100 4801 Rev FA October 2010 4 16 Rosemount 3051S Series ...
Page 98: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series 5 8 ...
Page 194: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series A 46 ...
Page 205: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 11 Rosemount 3051S Series ...
Page 206: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 12 ...
Page 207: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 13 Rosemount 3051S Series ...
Page 208: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 14 ...
Page 209: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 15 Rosemount 3051S Series ...
Page 210: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 16 ...
Page 211: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 17 Rosemount 3051S Series ...
Page 212: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 18 ...
Page 213: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 19 Rosemount 3051S Series ...
Page 214: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 20 ...
Page 215: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 21 Rosemount 3051S Series ...
Page 216: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 22 ...
Page 217: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 23 Rosemount 3051S Series ...
Page 218: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 24 ...
Page 219: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 25 Rosemount 3051S Series ...
Page 221: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 27 Rosemount 3051S Series ...
Page 222: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 28 ...
Page 223: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 29 Rosemount 3051S Series ...
Page 224: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 30 ...
Page 225: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 31 Rosemount 3051S Series ...
Page 226: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 32 ...
Page 227: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 33 Rosemount 3051S Series ...
Page 228: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 34 ...
Page 229: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 35 Rosemount 3051S Series ...
Page 230: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 36 ...
Page 231: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 37 Rosemount 3051S Series ...
Page 232: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 38 ...
Page 233: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 39 Rosemount 3051S Series KEMA ...
Page 234: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 40 ...
Page 235: ...Reference Manual 00809 0100 4801 Rev FA October 2010 B 41 Rosemount 3051S Series ...
Page 236: ...Reference Manual 00809 0100 4801 Rev FA October 2010 Rosemount 3051S Series B 42 ...
Page 239: ......