52
Reference Manual
00809-0100-2240, Rev DA
October 2017
Insert distance
You can specify a minimum distance between the product surface and the first temperature spot
element to be included in the average temperature calculation. If the temperature spot element is within
or above the Insert Distance, the element will be excluded from the calculation.
Figure 5-2. Insert distance
This function may be useful if the temperature of the atmosphere above the product surface
significantly deviates from the temperature of the product itself, resulting in large temperature
gradients close to the product surface. By specifying an Insert Distance, temperature elements within
this region can be excluded from average temperature calculations.
The Insert Distance function can also be used to compensate for inaccuracies in the measured
temperature element positions, in order to make sure that elements above the product surface are not
included in the average temperature calculation. If, for example, temperature element positions are
measured with an accuracy of 10 mm, setting the minimum distance to at least 10 mm will guarantee
that sensors above the surface are not included in the average temperature calculations.
Insert Distance
These sensors are not
included in the average
temperature
calculation
Product surface
Summary of Contents for Rosemount 2240S
Page 2: ......
Page 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Page 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Page 193: ......