119
July 2017
Reference Manual
00809-0100-2230, Rev CA
B.5 Europe
I1
ATEX Intrinsic Safety
Certificate: FM10ATEX0046X
Standards: EN 60079-0:2012 + A11:2013,
EN 60079-11:2012,
EN 60529:2013
Markings:
II 1 G Ex ia IIC Ga T4 Ta = -50°C to +70°C;
Entity/FISCO; IP66, IP67
Specific Conditions for Safe Use (X):
1. The non-metallic parts incorporated in the enclosure of this
equipment may generate an ignition-capable level of
electrostatic charge. Therefore particularly when it is used for
applications that specifically require Division 1 and Group II,
Zone 0 located equipment, the equipment shall not be
installed in a location where the external conditions are
conducive to the build-up of electrostatic charge on such
surfaces. Additionally, the equipment shall only be cleaned
with a damp cloth.
2. The enclosure contains aluminum and is considered to
present a potential risk of ignition by impact or friction. Care
must be taken during installation and use to prevent impact
or friction.
3. The Rosemount 2230 Graphical Field Display will not pass the
500Vrms dielectric strength test and this must be taken into
account during installation.
4. To maintain ingress protection ratings IP66 and IP67, PTFE
tape or pipe dope is required for cable entries and blanking
plugs.
B.6 International
I7
IECEx Intrinsic Safety
Certificate: IECEx FMG 10.0021X
Standards: IEC 60079-0:2011, IEC 60079-11:2011
Markings: Ex ia IIC Ga T4 (-50°C
Ta
+70°C); Entity/FISCO;
IP66/IP67
Specific Conditions for Safe Use (X):
1. The non-metallic parts incorporated in the enclosure of this
equipment may generate an ignition-capable level of
electrostatic charge. Therefore particularly when it is used for
applications that specifically require Division 1 and Group II,
Zone 0 located equipment, the equipment shall not be
installed in a location where the external conditions are
conducive to the build-up of electrostatic charge on such
surfaces. Additionally, the equipment shall only be cleaned
with a damp cloth.
2. The enclosure contains aluminum and is considered to
present a potential risk of ignition by impact or friction. Care
must be taken during installation and use to prevent impact
or friction.
3. The Rosemount 2230 Graphical Field Display will not pass the
500Vrms dielectric strength test and this must be taken into
account during installation.
4. To maintain ingress protection ratings IP66 and IP67, PTFE
tape or pipe dope is required for cable entries and blanking
plugs.
Ui
Ii
Pi
Ci
Li
Entity parameters
30 V
300 mA
1.3 W
2.1 nF
1.1 μH
FISCO parameters
17.5V
380 mA
5.32 W
N/A
N/A
Ui
Ii
Pi
Ci
Li
Entity parameters
30 V
300 mA
1.3 W
2.1 nF
1.1 μH
FISCO parameters
17.5V
380 mA
5.32 W
N/A
N/A
Summary of Contents for Rosemount 2230
Page 1: ...Reference Manual 00809 0100 2230 Rev CA July 2017 Rosemount 2230 Graphical Field Display ...
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 2230 Rev CA Title Page July 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2230 Rev CA Introduction July 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2230 Rev CA Overview July 2017 Overview ...
Page 44: ...34 Reference Manual 00809 0100 2230 Rev CA Installation July 2017 Installation ...
Page 154: ...144 Reference Manual 00809 0100 2230 Rev CA Index July 2017 Index ...
Page 155: ......