32
Reference Manual
00809-0100-4107, Rev BA
Section 2: Configure
April 2013
Changing transmitter address using AMS Device Manager
Right click on the device and select
Configure
.
1.
In HART Revision 5 mode:
a.
Click on
Manual Setup
, select the
HART
tab.
b. In the Communication Settings box enter polling address in the
Polling Address
box,
click
Send
.
2. In HART Revision 7 mode:
a.
Click on
Manual Setup
, select the
HART
tab and click the
Change Polling Address
button.
3. Carefully read the warning and click
Yes
if it is safe to apply the changes.
2.11.2
Communicating with a multidropped transmitter
To communicate with a multidrop transmitter, the Field Communicator or AMS Device Manager
has to be set up for Polling.
Communicating with a multidropped transmitter using a Field
Communicator
1.
Select
Utility
and
Configure HART Application
.
2. Select
Polling Addresses.
3. Enter
0-63.
Communicating with a multidropped transmitter using AMS
Device Manager
Click on the
HART modem icon
and select
Scan All Devices
.