Table 7-1: Echo Peak Types
Type
Description
Surface
Echo tracked as the current surface echo
Unknown
Echo identified as unknown (can be a surface candidate)
Suppressed
Echoes that are identified but suppressed by the device
Tank bottom echo
Echo considered as an echo from the tank bottom
7.5.2
Perform a device reset
The function is used to reset/restart the electronics without re-cycling the power.
Procedure
1. Under
Menu
, select
Parameter
→
Service Tools
→
Maintenance
.
2. Select
Device Reset
.
7.5.3
Restore to factory settings
This function restores the transmitter to factory settings (user configuration is
overwritten).
Procedure
1. Under
Menu
, select
Parameter
→
Service Tools
→
Maintenance
.
2. Select
Restore Factory Settings
.
7.5.4
Use the simulation mode
It is possible to set the level to a simulated value for testing purposes. The simulated value
affects both digital and analog output.
Procedure
1. Under
Menu
, select
Parameter
→
Service Tools
→
Simulation
.
2. In the
Simulated Level
box, enter the desired value.
3. Select
Write to device
.
4. Select
Start simulation (60 min)
.
7.5.5
Calibrate analog out
Use this function to calibrate the analog output by comparing the actual output current
with the nominal 4 mA and 20 mA currents. Calibration is done at factory and the analog
output does not normally need to be recalibrated.
Prerequisites
Connect a calibrated ampere meter to the analog output loop.
Service and troubleshooting
Reference Manual
September 2020
00809-0100-4480
66
Rosemount 1408H Level Transmitter