5.5
Perform the basic setup
5.5.1
Set the engineering units
Procedure
1. Under
Menu
, select
Parameter
→
Basic Setup
.
2. In the
Engineering Units
list, select
Metric
or
Imperial
.
Option
Description
Metric
Length: m
Temperature: °C
Volume flow: m
3
/h
Imperial
Length: in.
Temperature: °F
Volume flow: US gal/h
3. Select
Write to device
.
5.5.2
Enter the reference height
Procedure
1. Under
Menu
, select
Parameter
→
Basic Setup
.
2. Enter the Reference Height.
3. Select
Write to device
.
Related information
5.5.3
Configure the analog output
The transmitter can be set to output the level or volume flow as a 4-20 mA signal.
Procedure
1. Under
Menu
, select
Parameter
→
OUT2 Analog Output
.
2. In the
OUT2 Configuration
list, select
Analog Output 4-20 mA
.
3. In the
Analog Control Variable
list, select
Level
or
Volume Flow
.
4. In the
Alarm Mode
list, select
Low Alarm
or
High Alarm
.
5. Select
Analog Range Values
, and then enter the desired Upper Range Value (20
mA) and Lower Range Value (4 mA).
6. Select
Write to device
.
Reference Manual
Configuration
00809-0200-4480
May 2021
Reference Manual
33