CAUTION! This product uses semiconductors which can be damaged by electrostatic
discharge (ESD). When handling the printed circuit boards (PCBs), observe proper ESD
precautions so that the electronics are not damaged.
Electrical Installation
Installation • 9
3.4.2. Power and Signal Wiring
The product comes with cable glands and plugs pre-installed. The power entry cable
gland is without a gland plug. Use the appropriate cable glands to insert and connect
the wires for power and signal to the appropriate terminals as indicated in the figure
and wiring table that follow. The PCB terminal blocks are a pluggable type and may
be removed to aid termination.
IMPORTANT:
• For 24VAC installations sharing a transformer in a daisy-chain configuration, neutral
polarity must be maintained for all instruments.
• 24VAC power polarity must not be reversed.
• For a more robust system, a dedicated transformer for each MRLDS is recommended to
prevent damage caused by wiring errors.
• Fasten terminal screws.
Table 3-1
- Power and Signal Wiring
Power
Description
Label
Wiring Termination
Power
24VDC/VAC IN
24V IN: -
24VDC/VAC neutral/ground
24V IN: +
24VAC neutral
24VDC/VAC OUT
(power daisy chain
terminal)
24V OUT: -
24VDC/VAC neutral/ground
24V OUT: +
24VAC neutral
Digital Output
MODBUS Network
Communications
MODBUS: B
RS-485 “B” (inverted)
MODBUS: A
RS-485 “A” (non-inverted)
MODBUS: GND
RS-485 GND
MODBUS: SH
RS-485 Shield
Voltage or Current
ANALOG: -
Analog output ground
Analog
Output
Output
ANALOG: +
Analog output signal (+)
3.4.3. Relay Wiring
WARNING! Relays are rated for 0 to 30V AC/DC. DO NOT apply main power onto these
relays.