Ambient and process temperature limits for all models
122 (50)
D
A
B
C
125.5
(52)
140 (60)
-40 (-40)
-60 (–51)
-20 (-29)
20 (7)
60 (16)
100 (38)
-60 (-51)
300 (149)
-20 (-29)
20 (7)
60 (16)
100
(38)
140
(60)
200
(93)
280 (138)
240 (116)
320 (160)
A. Ambient temperature of core processor or transmitter in °F (°C)
B. Maximum process temperature in °F (°C)
C. Mount transmitter remotely and use a junction box
D. Temperature is below -40 °F (-40 °C); mount transmitter remotely and use a junction box
2.4
Recommendations for hygienic and self-
draining applications
For optimal cleanability and drainability:
•
If possible, install the sensor in a vertical pipeline with the process fluid flowing upward
through the sensor.
•
The minimum angle of inclination of the sensor depends on the process fitting. See
.
•
For clean-in-place (CIP) applications, Micro Motion recommends using the generally-
accepted flow velocity of at least 1.5 m/s for cleaning the sensor.
•
If the process piping must be larger than the sensor, eccentric reducers may be used to
ensure full drainability. In this case, the process end connections for the piping and
sensor must be the same size. See
.
Note
As part of the cleaning process, you can purge skid-based systems with nitrogen at the
end of the cleaning cycle. When using eccentric reducers, it is possible to trap gas in
the section of process piping adjacent to the reducer. Sensor performance can be
impacted by intermittent flow of the captured gas in a liquid fluid stream.
Installation Manual
Planning
20002172
February 2020
Installation Manual
9
Summary of Contents for Micro Motion T Series
Page 4: ...Contents Installation Manual February 2020 20002172 iv Micro Motion T Series ...
Page 12: ...Planning Installation Manual February 2020 20002172 12 Micro Motion T Series ...
Page 16: ...Mounting Installation Manual February 2020 20002172 16 Micro Motion T Series ...
Page 24: ...Grounding Installation Manual February 2020 20002172 24 Micro Motion T Series ...
Page 27: ...Installation Manual 20002172 February 2020 Installation Manual 27 ...