Options for
Flow Direction
(continued)
Table 4-4:
Flow direction setting from
ProLink III
Relationship to Flow Direction arrow on sensor
Negate Bidirectional
Appropriate when both forward and reverse flow are ex-
pected, and reverse flow will dominate, but the amount of
forward flow will be significant.
Effect of
Flow Direction
on flow totals
Flow Direction
affects how flow totals and inventories are calculated.
Effect of the
Flow Direction
parameter and actual flow direction on flow
totals and inventories
Table 4-5:
Flow Direction
setting
Actual flow direction
Forward
Zero flow
Reverse
Forward
Totals increase
Totals do not change
Totals do not change
Reverse
Totals do not change
Totals do not change
Totals increase
Bidirectional
Totals increase
Totals do not change
Totals decrease
Absolute Value
Totals increase
Totals do not change
Totals increase
Negate Forward
Totals do not change
Totals do not change
Totals increase
Negate Bidirectional
Totals decrease
Totals do not change
Totals increase
4.5
Configure density measurement
The density measurement parameters control how density is measured and reported.
Density measurement (along with mass measurement) is used to determine liquid volume
flow.
4.5.1
Configure
Density Measurement Unit
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Overview
Density Measurement Unit
controls the measurement units that will be used in density
calculations and reporting.
Configure process measurement
30
Micro Motion
®
LNG Series Meters
Summary of Contents for Micro Motion LNG Series
Page 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Page 6: ...Contents iv Micro Motion LNG Series Meters ...
Page 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Page 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Page 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Page 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Page 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Page 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Page 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Page 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Page 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Page 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Page 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Page 99: ...Default values and ranges Configuration and Use Manual 93 ...