Configure process measurement
Table 5-11
API reference tables, associated process fluids, and associated calculation values
continued
Table name
Process fluid
CTL source data
Reference
temperature
Density unit
23A
Generalized crude
and JP4
Observed density
and observed
temperature
60 °F (non-
configurable)
Relative density
Range: 0.6110 to
1.0760
23B
Generalized
products
Observed density
and observed
temperature
60 °F (non-
configurable)
Relative density
Range: 0.6535 to
1.0760
23D
Lubricating oils
Observed density
and observed
temperature
60 °F (non-
configurable)
Relative density
Range: 8520 to
1.1640
24C
Liquids with a
constant density
base or known
thermal expansion
coefficient
User-supplied
reference density (or
thermal expansion
coefficient)
and observed
temperature
60 °F (non-
configurable)
Relative density
53A
Generalized crude
and JP4
Observed density
and observed
temperature
15 °C (configurable)
Base density
Range: 610 to 1075
kg/m
3
53B
Generalized
products
Observed density
and observed
temperature
15 °C (configurable)
Base density
Range: 653 to 1075
kg/m
3
53D
Lubricating oils
Observed density
and observed
temperature
15 °C (configurable)
Base density
Range: 825 to 1164
kg/m
3
54C
Liquids with a
constant density
base or known
thermal expansion
coefficient
User-supplied
reference density (or
thermal expansion
coefficient)
and observed
temperature
15 °C (configurable)
Base density in kg/m
3
5.10
Configure the concentration measurement application
Display
Not available
ProLink II
ProLink→Configuration→CM Setup
Field Communicator
Configure→Manual Setup→Measurements→Concentration Measurement
The concentration measurement parameters control how the transmitter calculates concentration
from temperature and density data.
Configuration and Use Manual
93