Configure device options and preferences
Tip
Record your password for future reference.
6.4
Configure the speed of the transmitter’s response to
changes in process data
Display
Not available
ProLink II
ProLink→Configuration→Device→Response Time
Field Communicator
Not available
You can configure the speed of the transmitter’s response to changes in process data.
The following parameter is used to control the speed of the transmitter’s response:
•
Response Time
6.4.1
Configure
Response Time
Display
Not available
ProLink II
ProLink→Configuration→Device→Response Time
Field Communicator
Not available
Response Time
controls the rate at which the transmitter updates its outputs to reflect changes in process
data.
Procedure
Set
Response Time
as desired.
Option
Description
Normal
(default)
Transmitter outputs track process data at the
standard speed.
Special
Transmitter outputs track process data as fast as
possible.
Tip
If you set
Response Time
to
Special
, additional process “noise” will be present in the transmitter outputs.