Installation Manual

19

Wiring

Figure 3-8: Terminal block for sensor wiring on the DIN rail

Figure 3-9: 4-wire cable to Model 3500 with I/O cables for standard and remote core 
processors

2 11/16

(68)

TS 15

TS 32

TS 35 x 7,5

TS 35 x 15

9

(229)

3 1/16

(77)

2 1/4

(57)

2 5/16

(58)

2 1/2

(63)

2 9/16

(66)

inches

(mm)

Terminal block for sensor wiring

Cable for sensor wiring 

(attached to Model 3500)

Connector

Core processor terminals

4-wire cable

Maximum cable length:

See 

Table 1-1

VDC+

(Red)

VDC–

(Black)

RS-485B

(Green)

RS-485A

(White)

User-supplied or 

factory-supplied cable

VDC+

(Red)

RS-485A

(White)

VDC–

(Black)

RS-485B

(Green)

 Model 3500 

Terminal block for sensor wiring installed 

on DIN rail

Summary of Contents for Micro Motion 3300

Page 1: ...Installation Manual 20000885 Rev BA February 2015 Micro Motion Model 3500 Transmitter MVD or Model 3300 Controller Installation manual for panel mount ...

Page 2: ...en returning equipment These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees Failure to follow Micro Motion procedures will result in your equipment being refused delivery Information on return procedures and forms are available on our web support system at www micromotion com or by phoning the Micro ...

Page 3: ...gths 8 1 4 Install the Model 3300 or Model 3500 in the panel 9 Chapter 2 Mounting 11 2 1 Mount the core processor 11 Chapter 3 Wiring 13 3 1 Connect the input and output wiring 13 3 2 Connect the Model 3500 to the sensor 16 3 3 Wire the sensor to the remote core processor 24 3 4 Connect the power supply wiring 25 ...

Page 4: ...4 Model 3500 Transmitters or Model 3300 Peripherals Panel Mount ...

Page 5: ... 5 7 16 inch 138 mm square cutout in a panel that is 5 64 inch 2 mm to 1 2 inch 13 mm thick The bezel provides a IP65 watertight seal between the panel cutout and the applications platform housing See Figure 1 1 In addition the installation kit includes either A bracketed wiring connector for screw type connectors See Figure 3 2 I O cables and connectors See Figure 3 3 WARNING Improper installatio...

Page 6: ...l 3300 or Model 3500 where the ambient temperature is between 4 to 140 F 20 to 60 C 1 2 2 Dimensions See Figure 3 1 and Figure 3 3 for illustrations of screw type connectors versus I O cables Model type Dimension drawing Screw type wiring connectors Figure 1 2 I O cables Figure 1 3 Power supply wiring connector Retaining clip assembly for power supply wiring 2 x mounting bracket with screw Frame B...

Page 7: ...ns with I O cables 4 1 2 114 6 1 2 165 Bezel Panel 6 152 Approvals tag Frame Intrinsic safety shield Model 3500 only 7 3 4 197 1 2 12 8 1 2 216 6 152 inches mm 6 11 16 170 6 1 2 165 Bezel Panel 6 152 6 11 16 170 Minimum 4 1 4 108 mm bend radius Approvals tag 1 2 12 8 3 4 222 5 5 16 135 Frame 6 152 inches mm ...

Page 8: ...d the Model 3300 frequency input is 500 feet 150 meters Figure 1 4 4 wire remote transmitter Figure 1 5 Remote core processor with remote transmitter Installation type Maximum cable length 4 wire remote transmitter See Figure 1 4 and Table 1 1 for maximum length of the 4 wire cable Remote core processor with remote transmitter See Figure 1 5 and Table 1 1 for maximum length of the 4 wire cable and...

Page 9: ... watertight seal between the gasket and the panel Figure 1 6 Panel installation steps Table 1 1 Maximum cable lengths between sensor and transmitter Cable type Wire gauge Maximum length Micro Motion 4 wire Not applicable 1000 ft 300 m without Ex approval 500 ft 150 m with IIC rated sensors 1000 ft 300 m with IIB rated sensors Micro Motion 9 wire Not applicable 60 ft 20 m User supplied 4 wire VDC 2...

Page 10: ...10 Model 3500 Transmitters or Model 3300 Peripherals Panel Mount Planning ...

Page 11: ...tion go to Section 3 1 Figure 2 1 shows both types of core processors and mounting bracket Using the mounting bracket mount the core processor in a location compatible with the cable length requirements discussed in Section 1 2 Figure 2 1 Remote core processor and remote enhanced core processor components Core processor lid Upper conduit ring Core processor Lower conduit ring End cap Mounting brac...

Page 12: ...12 Model 3500 Transmitters or Model 3300 Peripherals Panel Mount Mounting ...

Page 13: ...iate terminals in the input output wiring connector Refer to the card that is inserted into the sleeve on the top panel shown in Figure 3 2 and Table 3 1 Use 24 to 16 AWG 0 25 to 1 5 mm2 twisted pair shielded wire Ground the shields at a single point only Figure 3 2 Input output wiring terminal card for screw type connectors Non intrinsically safe input output wiring terminal strip Intrinsically s...

Page 14: ...nnector to the I O terminal block 8 Connect input and output wiring to the appropriate terminals on the I O terminal block Table 3 1 Input output wiring terminals for screw type connectors Terminal number Designation c 2 a 2 Primary 4 20 mA output HART c 4 a 4 Secondary 4 20 mA output c 6 a 6 Frequency input c 8 a 8 Discrete input 1 c 10 a 10 Discrete input 2 c 12 a 12 Frequency output c 14 a 14 D...

Page 15: ... O cable shielding The cable connector does not connect the I O cable shielding to the chassis ground Figure 3 4 Terminal block installation steps for I O wiring on a DIN rail Figure 3 5 Input output wiring terminal label for I O cables DIN rail type TS 15 DIN rail type TS 32 DIN rail type TS 35 x 7 5 3 3 64 77 9 229 Label 3 17 32 90 Cable for input output wiring Connector I O terminal block DIN r...

Page 16: ...scribed in the sensor documentation 3 To connect the cable to the transmitter a Identify the wires in the 4 wire cable Use the 4 wire cable supplied by Micro Motion This cable consists of one pair of 18 AWG 0 75 mm2 wires red and black for the VDC connection and one pair of 22 AWG 0 35 mm2 wire green and white for the RS 485 connection b For transmitters with screw type connectors connect the four...

Page 17: ...Motion Function Terminal I O cable c 4 4 Red VDC a 4 10 Black VDC c 6 7 White RS 485A a 6 3 Green RS 485B Core processor terminals 4 wire cable Maximum cable length See Table 1 1 VDC Red VDC Black RS 485B Green RS 485A White User supplied or factory supplied cable Model 3500 Sensor wiring terminals See Figure 3 1 VDC Red RS 485A White VDC Black RS 485B Green a a a a a a a a a a a a a a a a c c c c...

Page 18: ...rocessors Core processor terminals 4 wire cable Maximum cable length See Table 1 1 User supplied or factory supplied cable Model 3500 Sensor wiring terminals See Figure 3 1 VDC Red RS 485A White VDC Black RS 485B Green a a a a a a a a a a a a a a a a c c c c c c c c c c c c c c c c 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 VDC Red RS 485B Green RS 485A White VDC Black ...

Page 19: ... 15 9 229 3 1 16 77 2 1 4 57 2 5 16 58 2 1 2 63 2 9 16 66 inches mm Terminal block for sensor wiring Cable for sensor wiring attached to Model 3500 Connector Core processor terminals 4 wire cable Maximum cable length See Table 1 1 VDC Red VDC Black RS 485B Green RS 485A White User supplied or factory supplied cable VDC Red RS 485A White VDC Black RS 485B Green Model 3500 Terminal block for sensor ...

Page 20: ...y shielded heat shrink as described in Step 6 The shielded heat shrink provides a shield termination Installation method Procedure Unshielded wiring in continuous metallic conduit that provides 360 termination shielding for the enclosed wiring Go to Step 8 User supplied cable gland with shielded cable or armored cable terminate the shields in the cable gland Terminate both the armored braid and th...

Page 21: ...d clamping insert 6 For connection at the core processor housing prepare shielded cable as follows for armored cable omit steps d e f and g a Strip 4 1 2 inches 114 mm of cable jacket b Remove the clear wrap that is inside the cable jacket and remove the filler material between the wires c Remove the foil shield that is around the insulated wires leaving 3 4 inch 19 mm of foil or braid and drain w...

Page 22: ...ng the cable apply heat 250 F or 120 C to shrink the tubing See Figure 3 13 Figure 3 13 Shielded heat shrink covering exposed drain wires g Position the gland clamping insert so that the interior end is flush with the heat shrink h Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1 8 inch 3 mm past the O ring See Figure 3 14 Figure 3 14 Folded cloth i Insta...

Page 23: ...n the core processor See Figure 3 16 Figure 3 16 Connect the four wires to the numbered slots 10 Connect the core processor housing internal ground screw if earth ground is required Earth ground is required if the core processor cannot be grounded via sensor piping and local codes require internal ground connections Do not connect shield drain wires to this terminal 11 Reinstall and tighten the co...

Page 24: ... opening d Connect the wires to the plugs supplied with the core processor e Insert the plugs into the sockets inside the lower conduit ring See Figure 3 17 Connection type Procedure Screw type connectors Connect the four wires from the core processor to the appropriate terminals on the transmitter See Table 3 3 and Figure 3 6 Do not expose bare wires Do not ground the shield braid or drain wires ...

Page 25: ... conduit ring Do not ground to the core processor s mounting screw Do not ground the cable at the sensor junction box Ground the cable braid on both ends by terminating it inside the cable glands CAUTION Make sure that the wires are not caught or pinched when you close the housing to reduce the risk of measurement error or meter failure CAUTION Do not install power supply wiring in the same cable ...

Page 26: ...ure that all ground wiring has less than 1 ohm impedance 3 Plug the power supply wiring connector into the power supply wiring terminals See Figure 3 18 4 Slide the retaining clip over the wiring then tighten the screw to hold the clip in place See Figure 1 1 5 Optional Install a user supplied switch in the power supply line In Europe install the switch close to the Model 3300 or Model 3500 in ord...

Page 27: ... Japan T 81 3 5769 6803 F 81 3 5769 6844 Micro Motion Asia Emerson Process Management 1 Pandan Crescent Singapore 128461 Republic of Singapore T 65 6777 8211 F 65 6770 8003 Micro Motion United Kingdom Emerson Process Management Limited Horsfield Way Bredbury Industrial Estate Stockport SK6 2SU U K T 44 0870 240 1978 F 44 0800 966 181 2015 Micro Motion Inc All rights reserved The Emerson logo is a ...