Interaction between Density Damping and Added Damping
In some circumstances, both
Density Damping
and
Added Damping
are applied to the reported
density value.
Density Damping
controls the rate of change in the density process variable.
Added Damping
controls the rate of change reported via the mA output. If
mA Output Process Variable
is set to
Density
, and both
Density Damping
and
Added Damping
are set to non-zero values, density
damping is applied first, and the added damping calculation is applied to the result of the
first calculation.
4.5.4
Configure Density Cutoff
Display
Not available
ProLink II
ProLink
>
Configuration
>
Density
>
Low Density Cutoff
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Density
>
Density Cutoff
Overview
Density Cutoff
specifies the lowest density value that will be reported as measured. All
density values below this cutoff will be reported as 0.
Procedure
Set
Density Cutoff
to the value you want to use.
The default value for
Density Cutoff
is 0.2 g/cm
3
. The range is 0.0 g/cm
3
to 0.5 g/cm
3
.
Effect of Density Cutoff on volume measurement
Density Cutoff
affects liquid volume measurement. If the density value goes below
Density
Cutoff
, the volume flow rate is reported as 0.
Density Cutoff
does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value
configured for
Standard Gas Density
.
4.6
Configure temperature measurement
The temperature measurement parameters control how temperature data from the
sensor is reported. Temperature data is used to compensate for the effect of temperature
on the sensor tubes during flow measurement .
The temperature measurement parameters include:
•
Temperature Measurement Unit
•
Temperature Damping
Configure process measurement
50
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......