The default value for
Slug High Limit
is 5.0 g/cm
3
. The range is 0.0 to 10.0 g/cm
3
.
3.
Set
Slug Duration
to the number of seconds that the transmitter will wait for a slug
flow condition to clear before performing the configured slug flow action.
The default value for
Slug Duration
is 0.0 seconds. The range is 0.0 to 60.0 seconds.
Slug flow detection and reporting
Slug flow is typically used as an indicator of two-phase flow (gas in a liquid process or liquid
in a gas process). Two-phase flow can cause a variety of process control issues. By
configuring the slug flow parameters appropriately for your application, you can detect
process conditions that require correction.
Tip
To decrease the occurrence of slug flow alarms, lower
Slug Low Limit
or raise
Slug High Limit
.
A slug flow condition occurs whenever the measured density goes below
Slug Low Limit
or
above
Slug High Limit
. If this occurs:
•
A slug flow alarm is posted to the active alarm log.
•
All outputs that are configured to represent flow rate hold their last “pre-slug flow”
value for the configured
Slug Duration
.
If the slug flow condition clears before
Slug Duration
expires:
•
Outputs that represent flow rate revert to reporting actual flow.
•
The slug flow alarm is deactivated, but remains in the active alarm log until it is
acknowledged.
If the slug flow condition does not clear before
Slug Duration
expires, the outputs that
represent flow rate report a flow rate of 0.
If
Slug Duration
is set to 0.0 seconds, the outputs that represent flow rate will report a flow
rate of 0 as soon as slug flow is detected.
4.5.3
Configure Density Damping
Display
Not available
ProLink II
ProLink
>
Configuration
>
Density
>
Density Damping
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Density
>
Density Damping
Configure process measurement
48
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......