11.6
Perform a D3 and D4 density calibration (T-
Series sensors only)
For T-Series sensors, the optional D3 and D4 calibration could improve the accuracy of the
density measurement if the density of your process fluid is less than 0.8 g/cm
3
or greater
than 1.2 g/cm
3
.
If you perform the D3 and D4 calibration, note the following:
•
Do not perform the D1 and D2 calibration.
•
Perform the D3 calibration if you have one calibrated fluid.
•
Perform both the D3 and D4 calibrations if you have two calibrated fluids (other
than air and water). The calibrations must be performed without interruption, in the
order shown. Make sure that you are prepared to complete the process without
interruption.
11.6.1
Perform a D3 or D3 and D4 density calibration using
ProLink II
Prerequisites
•
During density calibration, the sensor must be completely filled with the calibration
fluid, and flow through the sensor must be at the lowest rate allowed by your
application. This is usually accomplished by closing the shutoff valve downstream
from the sensor, then filling the sensor with the appropriate fluid.
•
For D3 density calibration, the D3 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
3
-
Minimum difference of 0.1 g/cm
3
between the density of the D3 fluid and the
density of water. The density of the D3 fluid may be either greater or less than
the density of water.
•
For D4 density calibration, the D4 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
3
-
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
-
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
•
Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
Procedure
See
Measurement support
Configuration and Use Manual
199
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......