1. Make a service port connection to your transmitter.
2. Choose
Device Tools
>
Configuration
>
Weights & Measures
.
3. Set
Software Security
to
Enabled
.
10.2.3
Switch between secured and unsecured mode using the
switching utility
Prerequisites
The switching utility must be installed on your PC. The switching utility is available from
the Micro Motion web site:
www.micromotion.com/softwaredownloads
.
Before switching to unsecured mode, ensure that you will be able to switch back to
secured mode. Because switching to unsecured mode requires breaking the physical seal,
switching back to secured mode may require a site visit from a certified Weights &
Measures inspector and reinstallation of the physical seal.
Ensure that the wiring is in place for a service port connection. You may need to break the
seal and remove the clamp.
Procedure
•
To switch from secured mode to unsecured mode:
1. Run the switching utility.
2. Select the COM port to use for the connection.
3. Click
Disable Custody Transfer
.
•
To switch from unsecured mode to secured mode:
1. Run the switching utility.
2. Select the COM port to use for the connection.
3. Click
Enable Custody Transfer
.
10.3
Clear Status Alarm A027: Security Breach
Status Alarm A027: Security Breach is posted if the transmitter is switched to unsecured
mode or if the transmitter detects that the core processor ID has changed.
1.
If the transmitter is in secured mode, switch to unsecured mode.
2.
Take any required actions related to the cause of the security breach.
3.
Switch to secured mode.
Operate the transmitter with the Weights & Measures application
170
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......