126
Micro Motion
®
Model 2400S Transmitters with Analog Outputs
Measurement Performance
Figure 10-13
D1 and D2 density calibration – Communicator
Enter density of D1 fluid
Density Calibration
Complete
message
Calibration in Progress
message
D1 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D1 fluid
Fill sensor with D2 fluid
D2 calibration
OK
1 Dens Pt 1
Perform Cal
OK
Home
Enter density of D2 fluid
Density Calibration
Complete
message
Calibration in Progress
message
OK
2 Dens Pt 2
Perform Cal
OK
Home
Done
On-Line Menu >
3 Diag/Service >
3 Calibration >
2 Density cal
3 Diag/Service >
3 Calibration >
2 Density cal