![Emerson FISHER FIELDVUE DLC3010 Instruction Manual Download Page 8](http://html.mh-extra.com/html/emerson/fisher-fieldvue-dlc3010/fisher-fieldvue-dlc3010_instruction-manual_102332008.webp)
Instruction Manual
D102748X012
DLC3010 Digital Level Controller
Introduction and Specifications
May 2018
8
Table 1‐1. DLC3010 Digital Level Controller Specifications (continued)
Minimum Differential Specific Gravity (continued)
See 249 sensor specifications for standard displacer
volumes and standard wall torque tubes. Standard
volume for 249C and 249CP sensors is
∼
980 cm
3
(60
in
3
), most others have standard volume of
∼
1640 cm
3
(100 in
3
).
Operating at 5% proportional band will degrade
accuracy by a factor of 20. Using a thin wall torque
tube, or doubling the displacer volume will each
roughly double the effective proportional band.
When proportional band of the system drops below
50%, changing displacer or torque tube should be
considered if high accuracy is a requirement.
Mounting Positions
Digital level controllers can be mounted right‐ or
left‐of‐displacer, as shown in figure 2‐5.
Instrument orientation is normally with the coupling
access door at the bottom, to provide proper
drainage of lever chamber and terminal
compartment, and to limit gravitational effect on the
lever assembly. If alternate drainage is provided by
user, and a small performance loss is acceptable, the
instrument could be mounted in 90 degree rotational
increments around the pilot shaft axis. The LCD meter
may be rotated in 90 degree increments to
accommodate this.
Construction Materials
Case and Cover:
Low‐copper aluminum alloy
Internal:
Plated steel, aluminum, and stainless steel;
encapsulated printed wiring boards; Neodymium Iron
Boron Magnets
Electrical Connections
Two 1/2‐14 NPT internal conduit connections; one on
bottom and one on back of terminal box. M20
adapters available.
Options
J
Heat insulator
J
Mountings for Masoneilan
t
,
Yamatake, and Foxboro
t
‐Eckhardt displacers
available
J
Level Signature Series Test (Performance
Validation Report) available (EMA only) for
instruments factory‐mounted on 249 sensor
J
Factory Calibration: available for instruments
factory‐mounted on 249 sensor, when application,
process temperature and density(s) are supplied
J
Device is compatible with user‐specified remote
indicator
Operating Limits
Process Temperature:
Ambient Temperature and Humidity:
See below
Conditions
Normal
Limits
(1,2)
Transport and
Storage Limits
Nominal
Reference
Ambient
Temperature
-40 to 80
_
C
(-40 to 176
_
F)
-40 to 85
_
C
(-40 to 185
_
F)
25
_
C
(77
_
F)
Ambient
Relative
Humidity
0 to 95%,
(non‐condensing)
0 to 95%,
(non‐condensing)
40%
Altitude Rating
Up to 2000 meters (6562 feet)
Weight
Less than 2.7 Kg (6 lb)
NOTE: Specialized instrument terms are defined in ANSI/ISA Standard 51.1 -
Process Instrument Terminology.
1. LCD meter may not be readable below -20
_
C (-4
_
F)
2. Contact your
or application engineer if temperatures exceeding these limits are required.
Table 1‐2. EMC Summary Results—Immunity
Port
Phenomenon
Basic Standard
Test Level
Performance
Criteria
(1)(2)
Enclosure
Electrostatic discharge (ESD)
IEC 61000‐4‐2
4 kV contact
8 kV air
A
Radiated EM field
IEC 61000‐4‐3
80 to 1000 MHz @ 10V/m with 1 kHz AM at 80%
1400 to 2000 MHz @ 3V/m with 1 kHz AM at 80%
2000 to 2700 MHz @ 1V/m with 1 kHz AM at 80%
A
Rated power frequency
magnetic field
IEC 61000‐4‐8
60 A/m at 50 Hz
A
I/O signal/control
Burst
IEC 61000‐4‐4
1 kV
A
Surge
IEC 61000‐4‐5
1 kV (line to ground only, each)
B
Conducted RF
IEC 61000‐4‐6
150 kHz to 80 MHz at 3 Vrms
A
Note: RTD wiring must be shorter than 3 meters (9.8 feet)
1. A = No degradation during testing. B = Temporary degradation during testing, but is self‐recovering. Specification limit = +/- 1% of span.
2. HART communication was considered as “not relevant to the process” and is used primarily for configuration, calibration, and diagnostic purposes.