Using AMS Device Manager to configure the meter
Section 4: Configuration
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
June 2017
3-9000-765 Rev H
d.
Click the
All Variables
tab to view a graphical display of the Primary, Secondary,
Third and Fourth Variables.
Figure 4-6 AMS Device Manager - Service Tools All Variables status indicators
8.
Click
OK
to return to the Overview page.
9.
Enable the Write Protect switch on the CPU Module to protect the meter’s
configuration.
10.
From the Overview window, click
Display Meter K-Factors
. K-Factors are a read-only
values calculated from the Full scale volumetric flow rate used with frequency outputs
and the Maximum frequency for frequency output.