6.5.2 Time Mode
In this mode, ultrasonics are applied to the sample for a specified period of time.
Ultrasonics will remain on until the timeout period has been reached unless a STOP
condition occurs.
The following table shows the parameters of On Time mode, along the default, max. and
min. values.
Entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters. (See
for further details).
Table 6.6
Time Mode Parameters
Parameter
Default
Max. Value
Min. Value
On Time
0.05 s
9.99 s
0.05 s
Amplitude
50%
100%
10%
Off Time
0.05 s
9.99 s
0.05 s
Ground Detect
0.05 s
0.99 s
0.00 s
NOTICE
You can return to the ready screen without saving any change by pressing the ESC key
at any time.
Table 6.7
Time Mode Operational Sequence
Step
Action
Reference
1
Turn on power, wait for the LCD to show ready and current
mode.
2
Press the Mode/Configuration key once to get into mode
selection. Current selected mode icon will blink.
Summary of Contents for Branson LPX
Page 4: ...iv 100 412 196 REV 06 ...
Page 6: ...vi 100 412 196 REV 06 ...
Page 8: ...viii 100 412 196 REV 06 ...
Page 10: ...x 100 412 196 REV 06 Table 7 2 System Trouble Analysis Chart 73 Table 7 3 Alarms Errors 76 ...
Page 24: ...14 100 412 196 REV 06 ...
Page 33: ...100 412 196 REV 06 23 Chapter 3 Delivery and Handling 3 1 Delivery and Handling 24 ...
Page 38: ...28 100 412 196 REV 06 4 1 4 Declaration of Conformity Figure 4 1 Declaration of Conformity ...
Page 76: ...66 100 412 196 REV 06 ...
Page 85: ...100 412 196 REV 06 75 Figure 7 2 LPX Power Supply Interconnect Diagram ...