50
100-214-294 REV. 05
6.5.5
Cleaning Items (Treating Samples) in Timed Sonics Mode
NOTICE
To stop ultrasonics at any time, press the Sonics key.
Table 6.19 Treating samples in Timed Sonics Mode
Step
Action
1
Turn Main Power on.
2
Press the On/Standby key to turn on the unit.
3
If necessary, degas the liquid. See
4
Set the amount of time you wish the items to be cleaned:
Use up/down keys to increase/decrease cycle time (hold for quick
increments/decrements).
5
Set the tank temperature:
• Press the Fn key until the Set Temperature icon flashes
• Press the Up/Down keys to alter the setting to the tank temperature you wish
to reach
• Press the Heat key once to activate heat. The Heat icon appears
NOTICE
Units can be switched between °F or °C using the Fn key. See Fn key
description on
6.5.1 CPXH Series Explanation of Controls
6
Set the ultrasonic power level:
• Press the Fn key until the Power Level icon flashes
• Press the Up key to select high power ultrasonics or press the Down key to
select low power ultrasonics
7
Press the Sonics key to activate ultrasonics.
8
Place the items into a basket, perforated tray, or beakers in a positioning
cover.
9
If using beakers or a solid tray, add cleaning solution to beakers or tray
to cover the items.
10
Slowly lower the tray or beakers into the tank. Do not allow items to
contact the tank bottom.
11
When items are clean, slowly remove them from the tank.
12
Rinse clean items with clean, warm water and dry, if necessary.
Summary of Contents for Branson CPX1800
Page 4: ...iv 100 214 294 REV 05 ...
Page 6: ...vi 100 214 294 REV 05 ...
Page 8: ...viii 100 214 294 REV 05 ...
Page 21: ...100 214 294 REV 05 11 Chapter 3 Delivery and Handling 3 1 Unpacking Your Unit 12 ...
Page 35: ...100 214 294 REV 05 27 Chapter 5 Installation and Setup 5 1 Installing Your Unit 28 ...
Page 76: ...72 100 214 294 REV 05 ...