Section 6: Service and Troubleshooting
December 2014
Service and Troubleshooting
Note
Before changing the Upper Null Zone, check entered limit and deadband values for the High
Level Alerts. High Level Alerts should not be placed in the Upper Null Zone.
To set the Upper Null Zone do the following:
1.
Identify desired Upper Null Zone using the Echo Curve plot:
a.
Start the Echo Curve reading, refer to
“Reading the Echo Curve” on page 83
.
b.
View the Echo Curve plot to find out if there are disturbing echoes close to the tank
top.
2.
Set the desired Upper Null Zone value:
a.
From the
Home
Screen, go to
Configure > Manual Setup > Level Setup > Probe >
Advanced Probe Options
.
b.
Under
Upper Null Zone
, type the desired value into the box and then click
Send
.
Figure 6-15. Upper Null Zone
Figure 6-16. Upper Null Zone Is Extended to Block Out Disturbances at the Top of the Tank
Upper Null Zone
Upper Reference Point
Measurement Range
3.0
5.0
Upper Null
Zone
Disturbance
Product Surface Peak
AM
P
LITUDE
DISTANCE
Summary of Contents for 3308A Rosemount 3308 Series
Page 2: ......
Page 6: ...vi Reference Manual 00809 0100 4308 Rev BA Title Page December 2014 ...
Page 12: ...vi Reference Manual 00809 0100 4308 Rev BA Table of Contents December 2014 Table of Contents ...
Page 14: ...2 Reference Manual 00809 0100 4308 Rev BA Section 1 Introduction December 2014 Introduction ...
Page 52: ...40 Reference Manual 00809 0100 4308 Rev BA Section 3 Installation December 2014 Installation ...
Page 80: ...68 Reference Manual 00809 0100 4308 Rev BA Section 5 Operation December 2014 Operation ...
Page 196: ...4 Reference Manual 00809 0100 4308 Rev BA Index December 2014 Index ...
Page 197: ......