background image

Quick Installation Guide

00825-0100-4842, Rev. BA
November 2010

Level

Rosemount 3100 Series Ultrasonic Level 
Transmitter External Temperature Sensor

Part No :  03100-0001-0001
For use with ultrasonic transmitters 3102 and 3105

Introduction

For transmitter installations where 
the integral temperature sensor is 
not representative of the air / medium 
temperature in the process, the use 
of the external temperature sensor 
improves accuracy.

Sensor Speci

fi

 cation

316 Stainless steel body with two 
locknuts, M8 thread, 6 mm maximum 
panel thickness. Sensor working 
temperature -40 to 85 °C, cable 
working temperature -20 to 80 °C.  
Cable 5 m long, single twisted pair.

Note :

When connected to a 3105 
intrinsically safe transmitter, this 
sensor is classi

fi

 ed as ‘simple 

apparatus’ according to EN 50020 
paragraph 5.4.
Temperature class T6, 
-20 °C < Ta < 80 °C

www.rosemount.com