5.2.4  Connecting the measuring electrode

a)  Place the measuring electrode into the appropriate opening of the cell 

b)  Using the triaxial cable, connect the measuring electrode to the "SENSOR" socket on the rear of the 

titra tor.

5.3  Inserting a burette

Current generation burettes are fitted with a RFID transponder chip on the holder (can be recognized by 
the small cover 

C

, represented in black). This chip is used for reading and writing properties such as 

titrate name, concentration or usable life.

a)  Install the burette with holder on the titrant bottle (see bu-

rette operating instructions).

b)  Turn the arrestment knob in the opposite direction to the 

arrow.

c)  Take the burette from the holder and push it onto the 

titra tor from the left hand-side, as illustrated:

The recesses on the driver arm 

A

 must be parallel to 

the slide-on part of the burette.

The piston rod 

B

 must be in the home position.

d)  Turn the arrestment knob in the direction of the arrow to 

secure the burette.

e)  Insert the dispensing tube into the next available opening 

in the titration stand 

D

, which is located coun ter clock-

wise from the electrode 

E

.

When rinsing burettes and tubes, place the tips of dispensing tubes into a waste container.

5.4  Internal magnetic stirrer

A magnetic stirrer is standard on all models and fitted at the top of the housing, underneath the pivoted 
titra tion arm. Both On/Off switching and the stirrer speed are controlled by the software. 

Volumetric Titrator

17

Installation (components)

Summary of Contents for Aquastar AQV21

Page 1: ...Volumetric Titrator AQUASTAR AQV21 Installation Information...

Page 2: ......

Page 3: ...e power supply 13 5 2 Titration stand 14 5 2 1 Installing the titration stand 14 5 2 2 Installing the solvent bottle 15 5 2 3 Installing the waste bottle 16 5 2 4 Connecting the measuring electrode 17...

Page 4: ......

Page 5: ...en connecting a printer The instruments can either be controlled by Touchscreen or by using LabX PC software The large color Touch screen allows intuitive control by the user and flexibility in its ad...

Page 6: ...ormation cor ro sion caused by the ingress of gases Risk of corrosion Always test the titration vessel for firm seating in the titration head If it falls off you could injure yourself if working with...

Page 7: ...uipment and optional accessories are listed in the appropriate operating instructions Description Order number Karl Fischer titrator Separate power supply 100 240 Volt 51105795 Power cable country spe...

Page 8: ...g cell 51107474 Magnetic stirrer bar 51191159 5 ml interchangeable burette set DV1005 Silicone tube 133 mm 51107480 Drying tube holder 23915 3 Drying tube with cover 23961 Seal titration stand drying...

Page 9: ...pcs 51105600 Brown glass bottle 1 l 71296 2 Flat seals minimum order quantity 5 pcs 23981 Screw top for bottles 23937 2 Draining tube Dispensing tube 51107481 Adapter for immediate draining 51105594 D...

Page 10: ...tle drying tube C Waste bottle drying tube D Solvent manager E Waste bottle F Solvent bottle G Titration arm H Titration beaker I Internal magnetic stirrer J On Off button K Indicator light LED L Touc...

Page 11: ...Power supply COM Balance XS analytical balance ETHERNET Network Link to LabX PC software via network PC PC connection via USB Link to LabX PC software via USB inter face USB1 Printer barcode reader me...

Page 12: ...ion A Solvent bottle dispensing tube B Burette dispensing tube C Measuring electrode D Stopper E Silicone tube of drying tube F Waste bottle draining tube G Three hole adapter Volumetric Titrator 12 T...

Page 13: ...tched on by pressing the On Off button The LED flashes as the system starts up and then remains permanently lit The titrator is switched off by pressing the On Off button again All current tasks are i...

Page 14: ...eaded ring into the center groove of the titration arm and then pressing down onto the magnetic stirrer Tight en the fastening screw c Place the three hole adapter NS stopper and connecting piece into...

Page 15: ...ight ness c Screw the other end of the dispensing tube with the adjust ing sleeve into the left hand opening of the three hole adapter at the titration stand d Fill a drying tube fitted with a molecul...

Page 16: ...e and press into the appropriate opening of the solvent manager f Using the silicone tube 170 mm connect the drying tube to the appropriate connection on the Solvent Man ag er g Using the silicone tub...

Page 17: ...to the arrow c Take the burette from the holder and push it onto the titra tor from the left hand side as illustrated The recesses on the driver arm A must be parallel to the slide on part of the bur...

Page 18: ...priate operating instructions Description Order number Titration software LabX light titration LabX pro titration Instrument licenses for titrators LabX Threaded ring 51107459 Cover with Three hole ad...

Page 19: ...581 Silicone tube 170 mm Drying tube with cover 23961 2 flat seals 23981 minimum order quantity 5 pcs 51105600 Silicone tube pressure tube to solvent manager 850 mm 51105581 Silicone tube 170 mm 51105...

Page 20: ...296 Flat seal minimum order quantity 5 pcs 23981 Screw top for bottles 23937 O ring 4 pcs for screw tops 51107496 Molecular sieve 250 g 71478 Adapter for bottles made by Merck Germany Fisher USA 23774...

Page 21: ...100 23951 23952 23954 Visco Spoon 51107668 Syringes 100 pcs 1 ml 10 ml 71492 71482 Injection needle 80 x 0 8 mm for 1 ml injections 100 pcs 80 x 1 2 mm for 10 ml injections 12 pcs 71484 71483 Interch...

Page 22: ...OLEDO balance LC interface 229065 Connection cable for SARTORIUS balances RS9 RS25 51190363 Barcode reader with USB interface 21901297 USB cable to barcode reader 21901309 USB P25 compact printer incl...

Page 23: ...Protective cover for touchscreen 51105567 Volumetric Titrator 23 Optional accessories...

Page 24: ...ver sheet PET Protective cover Copolymer Chassis Stainless steel Titration stand Crastin PBT Lid titration beaker Polypropylene Threaded ring Polypropylene Three hole adapter Polypropylene Dispensing...

Page 25: ...10 Current Max 500 mA Stirrer pump Socket 5 pin mini DIN Pump detection Yes Stirrer detection Yes Pump voltage 24 V DC 5 max 400 mA Stirrer voltage range 0 to 18 V DC 10 max 300 mA SENSOR Socket LEMO...

Page 26: ...The devices have a power management system which prevents the titrator from switching off unex pect ed ly in the event of a power overload Tasks which would cause a power overload because a number of...

Page 27: ...s PG S COM interfaces XP COM interfaces XS COM interfaces Sartorius Various COM interfaces Standard Option Printer Printers can be connected to the USB1 or USB2 interface of the titrator PCL compatibl...

Page 28: ...r can also be used to enter text in open input fields Barcode readers can be connected to the USB1 or USB2 interface of the titrator A standard USB hub can be used if more than two devices are to be c...

Page 29: ...rting the titrator Note the following instructions when transporting the titrator to a new location Empty the titration beaker and remove it from the titration stand Remove all burettes and tube conne...

Page 30: ...code Mettler Toledo AG Analytical will retain possession of all documents and their reproductions and may wish to conclude a nondisclosure agreement with those requesting access to these documents Sch...

Page 31: ...10 Product life cycle model Volumetric Titrator 31 Product life cycle model...

Page 32: ......

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ...ibbstown NJ08027 Phone 856 423 6300 Fax 856 423 4389 Internet www emdchemicals com ME 51710714 0812 6 12 For AQUASTAR titrator Troubleshooting and Technical Support call 1 888 871 3535 For a listing o...

Reviews: