![EMC-PARTNER EXTTRA3000 E User Manual Download Page 129](http://html1.mh-extra.com/html/emc-partner/exttra3000-e/exttra3000-e_user-manual_2399826129.webp)
EMC PARTNER AG
129/162
12.3 Remote Control Command set
Command TST
(TeST)
Explanation:
set or query the test mode.
Arguments:
characters
IMP1, IMP2, IMP3..........., IMP11
Example:
TST
IMP1
This command must be used at a generator with different wave shapes.
Command VNOM
(Voltage NOMinal)
Set or query V-peak [in V]
Argument:
Integer
0..Vmax resp. 0..110 of DIP
Example:
VNOM
1500
VNOM?
Answer:
1500
Command POL
(POLarity)
Explanation:
Set or query the Polarity.
Argument:
Characters
POS, NEG
Example:
VNOM
1500
POL NEG
Command REP
(REPetition)
Explanation:
depends on the type of test:
Argument:
Integer
Example:
VNOM
1500
POL
NEG
REP 10
Command NBR
(NumBeR)
Explanation:
depends on the type of test:
Argument:
Integer
0..30000
Example:
NBR
10