7.4 Disassembly and Storage of the BioProTT
TM
Clamp-On SL
In order to not impair the measurement, it is best to disassemble the BioProTT
TM
Clamp-On SL
once the measurement process has been completed.
To disconnect the BioProTT
TM
Clamp-On SL, follow these steps:
1.
Open the spring-loaded lock and swing back the lid.
2.
Remove the tube from the channel.
3.
Close the lid and pull the plug out of the evaluation device.
Please note:
Only pull lightly on the sleeve of the sensor plug and make sure not to twist so the pins are not damaged.
Once you have disconnected the BioProTT
TM
Clamp-On SL, check for any residue of the medium or any other
kind of dirt or grime and clean the sensor if necessary. Store the sensor according to storage specifications.
Make sure to store it in a place where it cannot fall off or be dropped and ensure that the plug is dry.
7.5 Exchanging the BioProTT
TM
Clamp-On SL
If the BioProTT
TM
Clamp-On SL was exchanged before switching on the evaluation device, the new calibration
parameters will be shown on the display of the BioProTT
TM
FlowTrack SL after the next start-up sequence. If
the BioProTT
TM
Clamp-On SL is exchanged during operation, the alarm is activated with an error message.
Once the new BioProTT
TM
Clamp-On SL is connected, the alarm is reset and the calibration data can be seen
on the display.
•
Make sure that the BioProTT
TM
Clamp-On SL is attached at a different place on the tube each
time to avoid deformation.
•
Ensure that the calibration table used is the one needed for your respective application.
8
Adjustment and Calibration of the BioProTT
TM
Clamp-On SL
In order to ensure the highest possible accuracy, all BioProTT
TM
Clamp-On SLs are adjusted according to
customer specifications in regard to the relevant paramters prior to shipment.
If any of these parameters change, it is possible to send the sensors in for a re-adjustment.
In addition, em-tec GmbH recommends regular calibrations to be carried out. A rule of thumb would be every
24 months. Other than the adjustment, the calibration can be carried out by the customer on site.
For more information regarding this process, you can read and download our TechNote "Adjustment vs.
Calibration" on our website. It is, of course, also possible to send the sensor to em-tec GmbH so we can carry
out the calibation.