em-tec BioProTT Clamp-On SL User Manual Download Page 10

EN

Use of the BioProTT

TM

 Clamp-On SL

6.1 

General Safety Information 

• 

Inspect your BioProTT

TM

 Clamp-On SL for completeness and damage when unpacking it for the 

first time. 

• 

If the BioProTT

TM

 Clamp-On SL is not used as intended, the user may be exposed to risks 

that were not taken into account during its development.

• 

If the BioProTT

TM

 FlowMeasurement System is used to control the liquid volume flow, the 

user must analyze the risk of the application and, if necessary, take actions independent of 
the BioProTT

TM

 FlowMeasurement System to minimize the risk.  

• 

For the installation, and later the operation, all ambient conditions must meet the 

prescribed specifications (see 

chapter 10 “Technical Specifications”

).

• 

The user (= the person who connects the BioProTT

TM

 Clamp-On SL to the 

tubing circuit and measurement system) is responsible for any risk if the 

 

BioProTT

TM

 Clamp-On SL is not connected correctly.

• 

Do not drop the BioProTT

TM

 Clamp-On SL. It might break or get damaged, which could impair 

the measurement. 

6.2  Connecting the BioProTT

TM

 Clamp-On SL to the Evaluation Device

Lab Applications using the BioProTT

TM

 FlowTrack SL

To connect the BioProTT

TM

 Clamp-On SL to the BioProTT

TM

 FlowTrack SL, follow the steps listed below: :

  Carefully attach the round sensor plug to the sensor connection of the evalution device. 

For the BioProTT

TM

 FlowTrack SL, this is located at the back of the device.  

  If it is possible and necessary to use an extension cable, connect the sensor to one end of the extension 

cable and the other end of the extension cable to the evaluation device. 

  Once this is done, the sensor can be attached to the tube.

Industrial Applications using a device of the BioProTT

TM

 FlowMCP Series:

To connect the BioProTT

TM

 Clamp-On SL to one of the evaluation devices of the BioProTT

TM

 FlowMCP Series, 

follow the steps listed below: 

  Carefully attach the extension cable to the respective BioProTT

TM 

FlowMCP (D-Sub). 

 

For the BioProTT

TM

 FlowMCP Series, this is located at the bottom of the respective device 

  Connect the sensor to other end of the extension cable (round plug). 

  Once this is done, the sensor can be attached to the tube.

Please note:

 

The total cable length between the respective evaluation device and the 

 

BioProTT

TM

 Clamp-On SL shall not exceed 4 m. 

BioProTT™ Clamp-On SL

Page  10

Copyright em-tec GmbH

D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 

DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75

Summary of Contents for BioProTT Clamp-On SL

Page 1: ...BioProTTTM Clamp On SL User Manual DocuSign Envelope ID BE88571C E493 4DFB AF9A 864EB3FC3D75...

Page 2: ...c GmbH Finning All rights reserved Subject to Technical Changes Owing to our policy of continuous product development the illustrations and technical data contained in this document may differ slightl...

Page 3: ...Maintenance and Service 9 6 Use of the BioProTTTM Clamp On SL 10 6 1 General Safety Information 10 6 2 Connecting the BioProTTTM Clamp On SL to the Evaluation Device 10 6 3 Attaching the BioProTTTM Cl...

Page 4: ...itoring and control of the upstream feed monitoring and control of the downstream feed filtration fill and finish processes process evaluation upscaling processes research The SL at the end of the pro...

Page 5: ...Clamp On SLs incorporate two pairs of transducers i e four piezo ceramics in total which are arranged in an X configuration around the tube containing the measured flow When sending ultrasonic signals...

Page 6: ...aning Do not dispose of this device together with domestic waste The device as a whole as well as any parts must be disposed of in accordance with WEEE Directive and national legislation Manufacturer...

Page 7: ...Clamp On SL is part of to ensure that the BioProTTTM Clamp On SL is used inspected and maintained in accordance with the user manual This also applies for all subsequent revisions or instructions from...

Page 8: ...8 Troubleshooting as well as to the user manual of the evaluation device respectively 5 5 Electrical Safety and Electromagnetic Compatibility The BioProTTTM FlowMeasurement System i e the BioProTTTM C...

Page 9: ...service for the BioProTTTM Clamp On SL may be carried out by em tec GmbH only If these instructions are not followed em tec GmbH shall accept no liability for the component and the warranty will be vo...

Page 10: ...TTM FlowTrack SL To connect the BioProTTTM Clamp On SL to the BioProTTTM FlowTrack SL follow the steps listed below Carefully attach the round sensor plug to the sensor connection of the evalution dev...

Page 11: ...t deformed damaged or dirty Ensure that the tube size and the tube material used corresponds with the size of the BioProTTTM Clamp On SL Air in the tube can lead to errors in the measurement or interr...

Page 12: ...ease refer to the user manual of the respective BioProTTTM evaluation device Make sure to follow the steps described below and to heed the warnings included in them in order to ensure the highest poss...

Page 13: ...ued or no flow value is displayed as the low coupling might impair the measurement It is important that any unwanted interruption of the process due to damage or kinking of the tubing is avoided at al...

Page 14: ...ction of the BioProTTTM Clamp On SL Make sure that no detergents or disinfectants leak into the sensor plug For the cleaning of the device we recommend commonly used detergents and or the recommended...

Page 15: ...ensor or to fix it in one position em tec offers optional mounting solutions The features differ slightly in regard to their design and handling depending on the sensor size Sensors with an OD of up t...

Page 16: ...or torsial moment caused by the tube are too high the thread may tear the sensor base body may break or the sensor might be levered out If the mounting feature and thus sensor is not positioned corre...

Page 17: ...mmends regular calibrations to be carried out Ideally the time between re calibrations should not exceed 24 months Other than the adjustment the calibration can be carried out by the customer on site...

Page 18: ...nfluence the sensor or if the sensor or its cable came into contact with an electric cautery 2 Check the grounding of the valves and pumps or any other devices that might create electromagnetic distur...

Page 19: ...nsor lid is closed properly 2 Close the lid If the problem persists contact our service department The tube size and or material is not the one indicated for the sensor 1 Check if the tube size and ma...

Page 20: ...hannel If the problem persists contact our service department There might be electromagnetic fields influencing the sensor 1 Check if the sensor is damaged or not connected properly to the device 2 Ch...

Page 21: ...sories must be disposed of in accordance with the applicable national provisions for electronic components In accordance with the requirements of EU Directive 2012 65 EC Waste Electrical and Electroni...

Page 22: ...to OD 7 8 2 of reading 80 ml min up to OD 1 1 8 2 of reading 200 ml min up to OD 1 5 8 2 of reading 400 ml min The specified accuracies were determined under the following conditions Medium Water Med...

Page 23: ...9 16 13433 PCT 3 8 x 1 8 0 017 2 5 0 100 10 3 8 1 8 5 8 13422 PCT 1 2 x 3 32 0 017 2 5 0 100 16 9 2 40 1 2 3 32 11 16 13428 PCT 1 2 x 1 8 0 017 2 5 0 100 16 9 2 40 1 2 1 8 3 4 13430 PCT 1 2 x 3 16 0 0...

Page 24: ...1 2 x 3 16 PCT 3 4 x 1 8 PCT 3 4 x 3 16 PCT 1 x 1 8 PCT 1 x 3 16 PCT 1 1 4 x 3 16 Size H x W x D 85 x 80 x 10 6 mm 100 x 80 x 10 6 mm Weight 28 g 31 g Material polyamide polyamide Mounting four M5 scr...

Page 25: ...nue of approximately 7 billion We deliver innovative equipment and components consumable supplies aftermarket parts software and digital solutions and support services through five operating segments...

Reviews: