Quick guide for TBA firmware V1.070

VI. Thermal protection. 

(figure 44)

 The “Temperature Exceeded” information 

will appear when the temperature of the chamber exceeds 100°C 
or thermal security measures of the heater are activated.
Solution to the problem: Check if the device has proper ventilation. 
Cancel the emergency message following instructions in point 10. 
If the alarm occurs again, contact the support service.

In order to enable further work, the reason for the alert should be 
eliminated, e.g. the Emergency STOP safety switch should be turned on,

the [F3] - [Failures] button should be pressed.

The display will show information about emergencies, e.g.  “Emergency 
STOP”. 

(figure 45)

In case of any of the emergencies, after their cancellation or termination, 

press the [F3] - „Failures” button to open the list of failures and cancel 
the alert by means of the [F2] - „Cancel” button.
 Failures can not be deleted if the emergency is not resolved, e.g. the 

emergency switch is on.

Press the [F2] - „Cancel” button to delete the emergency and then 
press the [F1] - „Exit” button.

The basic screen can be seen on the display 

(figure 46)

.

Press the [F1] - „Exit” button to leave the basic screen and allow for 

the further work of the device.

The battery tester is protected from power failure. During the 
discharge process current data are saved in a non-volatile memory 
every 5 minutes. If power failure and power return occurred during 
the test, the display will show the message in 

(figure 47)

.

Press the [F3] - [Yes] button to continue the process.

Checking the voltage of the connected battery. 

(figure 48)

If the value of the displayed voltage is consistent with the voltage of 

the tested battery, press the [F3] - „Next” button. 

Then the tester will display the parameters of the battery before the 
power outage and will be prepared to continue the test.

To cancel the further discharge process, press the [F1] - „Cancel” button.

figure 45

figure 46

figure 47

figure 44

10. Responding to alerts.

11. Protection from power outage .

figure 48

Summary of Contents for TBA 150

Page 1: ...berecorded and transferred from the tester by means of a USB flash drive Table of Contents 1 Quick guide page 2 2 Description of the device dashboard page 2 3 Connecting batteries to the tester page 3 4 Starting the tester in the Discharge mode page 4 5 Starting the tester in the 5h 10h 20h Discharge mode page 6 6 Discharge of a battery with the rated voltage other than the basic voltages determin...

Page 2: ...ate of the device Safety switch Enables immediate emergency cessation of the device operation ON Launches the discharge process OFF Temporarily halts the discharge process Adjuster Enablessettingthevalueofagivenparameter F1 F2 F3 function buttons Designed to select options on the display USB slot Enables transfer of data recorded during a test to a data carrier of pendrive type 1 2 3 4 5 6 7 8 9 T...

Page 3: ...0 V or 220V the type of the battery is determined Figure 2 presents an exemplary screenshot after attaching a 48V battery If a battery with another rated voltage is attached to the tester figure 3 further proceeding should be conducted in accordance with point 6 Discharge of a battery with the rated voltage other than the basic voltagesdeterminedforgiventypesofbatterytesters Otherbattery NOTE In a...

Page 4: ...xt screen Data logging screen figure 6 Usingthe F2 buttonplacethecursoron Datalogging andbymeans of the adjuster choose the number of memory bank where the data will berecordedduring thebatterytest Youcanselect onlytheempty banks which is a protection against erasure of data from previous recordings In order to avoid recording choose the No option Using the F2 button place the cursor on Period and...

Page 5: ...heset discharge time is over figure 10 Fans work for three minutes after the end of the discharge Blowing process enables faster cooling of the resistors When the blowing time is over information about reason for the end ofthedischargewill remainonthedisplay U ifthebatteryvoltageat the Uk link achieved the set value figure 11 or t if the set discharge time was over figure 12 Press the F1 End butto...

Page 6: ...ter set the correct number of tested cells this parameter is unavailable in the TBA 150 testers To proceed to the next screen choose Next by pressing the button F3 Data logging screen figure 15 Using the F2 button place the cursor on Data logging and by means of the adjuster choose the number of memory bank where the data will be recorded during the battery test You can select only empty bankswhic...

Page 7: ...rge time was over t figure 19 Blowingisonfor3minutesaftertheendofthedischarge Theblowing process allows for faster cooling of resistors When the blowing time is over the display will show the information on the discharge end reason U when the set final voltage of the Vk cell achieved the set value figure 20 or t when the set discharge time was over figure 21 Press the F1 End button to return to th...

Page 8: ...cethecursoron Uk andsetthefinalvoltage value of the battery by means of the adjuster Using the F2 button place the cursor on Time and set the discharge time by means of the adjuster To proceed to the next screen choose Next by pressing the button F3 Data logging screen figure 24 Using the F2 button place the cursor on Data logging and by means oftheadjusterchoosethenumberofmemorybankwherethedatawi...

Page 9: ...re 27 or when the set time was over figure 28 Blowingisonfor3minutesaftertheendofthedischarge Theblowing process allows for faster cooling of resistors When the blowing time is over the display will show the information on the discharge end reason U when the Uk battery voltage achieved the set value figure 29 or t when the set discharge time was over figure 30 Press the F1 End button to return to ...

Page 10: ...an contain 512 records Records are saved in the following time interval the period of record saving The frequency of record saving is set before the discharge process Transfer of data from the tester memory to a PC should be performed by using a USB memory stick For USB memory sticks with the capacity up to 2 GB select the FAT16 file system while formatting them For USB memory sticks with the capa...

Page 11: ...press the F3 OK button The Logging and USB Memory screen figure 36 allows for managing therecorderbanks Selectabankbymeansoftheadjuster Therecord of the selected bank is deleted by means of the F2 Delete button Thetesterisequippedwithfunctionswarningagainstawrongdecision figure 37 To permanently delete the content of the data Bank press the F1 Yes button When the Bank data deletion is over the cur...

Page 12: ...eryshouldbeconnectedinaccordance with the signs III Reverse connection of the battery after pause and restart of the discharge process figure 41 Ifthepauseoftheprocessoccurs bythe OFF buttonbeingpressed information about the end of the discharge process is displayed The battery can now be disconnected and connected again however its reverse connection will prevent the tester from further work Solu...

Page 13: ... be deleted if the emergency is not resolved e g the emergency switch is on Press the F2 Cancel button to delete the emergency and then press the F1 Exit button The basic screen can be seen on the display figure 46 Press the F1 Exit button to leave the basic screen and allow for the further work of the device The battery tester is protected from power failure During the discharge process current d...

Page 14: ...n allows you to review data from the last discharge test performed Thedataarestoredinthedevicememoryuntilthepowerisswitchedoff II Choosing the cell type figure 52 By means of the adjuster you can change the cell type Acid Alkaline If you choose Acid cells the device will assign voltage parameters of the tested cells in accordance with acid cells during the 5h 10h 20h Discharge If you choose Alkali...

Page 15: ... should be set again To set the date and time use the F2 button and the adjuster and confirm the setting by pressing the F3 OK button c Counter settings figure 56 This function allows you to set the mode of counting time during a battery test If the parameter is set to Backwards the battery test time counts down to 0 If the parameter is set to Forwards the battery test time counts to a set value T...

Reviews: