39
6.3.4 Limestone
NOTICE
Calibrate analyser with calcium carbonate of 12% C.
Temperature 1250 °C or higher.
Weight
300mg
6.3.5 Oil, Asphalt and Rubber
Temperature 1450 °C
Weight
80mg-100mg (fuel oil and other thick oils).
Up to 50mg (thin oil analysis)
6.3.6 Wood
Temperature 1300 °C
Weight:
350mg
Set the minimum analysis time to 50s. Set the comparator level to 20mV.