EK280
Application Manual
37
6.3
Status output (signal output)
Each of the outputs 1 to 4 (terminals DA1 to DA4) can be programmed as a status output (signal output).
6.3.1
Parameterizing by keyboard
i
The use of
enSuite
parameterizing software is strongly recommended for parameterizing a
status output (see below).
Parameterization with the keyboard on the EK280 is described below.
Setting up the output mode:
Move the cursor to the tab "
Serv.
" and subsequently via the following path to the mode for the
corresponding output
"
Md.O1
"
to
"
Md.O4
"
:
Serv.
Outputs
Output 1
Md.O1
Serv.
Outputs
Output 2
Md.O2
Serv.
Outputs
Output 3
Md.O3
Serv.
Outputs
Output 4
Md.O4
i
The status outputs of the EK280 can be operated with active or inactive logic. The output mode
must be programmed accordingly for this purpose (see below).
Press the ENTER
button. The actual parameterized value flashes.
Press one of the arrow buttons
or
until the
""
(Logic active: message is active, when the output is switched on.) or
"status-"
(Logic
inactive: message is active, when the output is switched off) flashes.
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.
Setting up the status display
for the output:
When in the corresponding menu press the
button repeatedly until
"SpO1
",
"SpO2
",
"SpO3
", or
"SpO4
" is displayed.
Press the ENTER
button. The actual parameterized value flashes.
Move the cursor to the digits using the
or
arrow keys and update them using the
or
arrow
keys to a value specified in the table.
Sc.O…
Description
0.mm_0s:1.1 *
A report in a Status
St.1
to
St.9
0.mm_02:2.1 *
A report in system status
StSy
0.mm_01:2.1 *
A report in overall status
Stat
1.mm_0s:1.1 *
Message group in system status
StSy
1.mm_02:2.1 *
VmD
Actual volume disturbance quantity
1.mm_01:2.1 *
Message group in overall status
Stat
* mm = report-number (1...16), s = status number (1...9 for St.1 ... St.9)
6.3.2
Parameterizing with enSuite
The setting of the required report (or range) in particular can be conveniently set with enSuite.
The paths which are described below can be found in
enSuite
under
"
Parameterize online
" > "
Parameters
" > "
Outputs
"
Setting up the output mode:
First of all, the mode must be set to the required value:
""
(Logic active: message is active, when the output is switched on.) or
"status-"
(Logic inactive: Report is active, when the output is switched off.)